We present ultrasensitive measurements of molecular absorption using frequency-agile rapid scanning, cavity ring-down spectroscopy with an external-cavity diode laser. A microwave source that drives an electro-optic phase modulator with a bandwidth of 20 GHz generates pairs of sidebands on the probe laser. The optical cavity provides for high sensitivity and filters the carrier and all but a single, selected sideband. Absorption spectra were acquired by stepping the tunable sideband from mode-to-mode of the ring-down cavity at a rate that was limited only by the cavity decay time. This approach allows for scanning rates of 8 kHz per cavity resonance, a minimum detectable absorption coefficient of 1.7 × 10~(-11) cm~(-1) after only 20 ms of averaging, and a noise-equivalent absorption coefficient of 1.7 × 10~(-12) cm~(-1) Hz~(-1/2). By comparison with cavity-enhanced laser absorption spectrometers reported in the literature, the present system is, to the best of our knowledge, among the most sensitive and has by far the highest spectrum scanning rate.
机译:我们提出了使用频率捷变快速扫描,外腔二极管激光器的腔衰荡光谱技术对分子吸收进行超灵敏测量。驱动带宽为20 GHz的电光相位调制器的微波源会在探测激光器上产生成对的边带。光学腔提供了高灵敏度,并过滤了载波和除单个选定边带以外的所有边带。通过将可调谐边带从振铃腔的各个模式步进到另一个模式来获取吸收光谱,该速率仅受腔衰减时间限制。这种方法允许每个腔谐振的扫描速率为8 kHz,平均仅20 ms后的最小可检测吸收系数为1.7×10〜(-11)cm〜(-1),等效噪声吸收系数为1.7×10 10〜(-12)厘米〜(-1)赫兹〜(-1/2)。通过与文献中报道的腔增强激光吸收光谱仪进行比较,就我们所知,本系统是最灵敏的光谱扫描速率,并且是迄今为止最高的光谱扫描速率。

著录项

  • 来源
    《Applied physics》 |2014年第4期|489-495|共7页
  • 作者单位

    National Institute of Standards and Technology, 100 Bureau Drive, Gaithersburg, MD 20899, USA;

    National Institute of Standards and Technology, 100 Bureau Drive, Gaithersburg, MD 20899, USA,Frequency Standards and Metrology Research Group, School of Physics, The University of Western Australia, Perth, WA 6009, Australia;

    National Institute of Standards and Technology, 100 Bureau Drive, Gaithersburg, MD 20899, USA;

    National Institute of Standards and Technology, 100 Bureau Drive, Gaithersburg, MD 20899, USA;

    National Institute of Standards and Technology, 100 Bureau Drive, Gaithersburg, MD 20899, USA;

  • 收录信息 美国《科学引文索引》(SCI);美国《工程索引》(EI);美国《生物学医学文摘》(MEDLINE);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号