The detection performance of a portable ~(214)Am ionization aspiration-type ion mobility spectrometer (M90-D1-C, Environics Oy) was investigated with nerve gases, blister agents, blood agents, choking agents and related compounds. The vapors of nerve gases, sarin, soman, tabun, cyclohexylsarin were recognized as "NERVE" after about several seconds of sampling, and the limits of detection (LOD) were <0.3 mg m~(-3) The vapors of blister agents, mustard gas and lewisite 1, and blood agents, hydrogen cyanide and cyanogen chloride were recognized as "BLISTER" with an LOD of <2.4 mg m~(-3) and >415 mg m~(-3), respectively. The vapor of chlorine was recognized as "BLOOD" with an LOD of 820 mg m . The vapors of nitrogen mustard 3 and chlorpicrin were recognized as different alarm classes, depending on their concentrations. The vapors of nitrogen mustard 1, 2 and phosgene did not show any alarm. As for interference, the vapors of nerve gas simulants, dimethylmethylphosphonate, trimethylphosphate, trieth-ylphosphate, diisopropylfluorophosphate, blister agent simulants, 2-chloroethylethylsulfide, 1,4-thioxane, 2-mercaptoethanol, and 20 organic solvents within 38 solvents examined were recognized false-positively. The patterns of detection sensor channel response values of 6 ion mobility cells and semiconductor cell were compared with the situation of the alarm against chemical-warfare agents.%化学剤(化学兵器用軌chemical warhre agent)は.即rn効性で,微量で殺傷性,無能力性を示し,神経ガス,びrnらん剤,窒息剤,血液剤,くしゃみ剤,催涙剤などに分類rnされが.
机译:研究了便携式〜(214)Am电离抽吸型离子迁移谱仪(M90-D1-C,Environics Oy)与神经毒气,起泡剂,血液制剂,窒息剂及相关化合物的检测性能。大约几秒钟的采样后,神经气体,沙林,梭曼,塔布恩,环己基沙蛋白的蒸气被识别为“神经”,并且检出限(LOD)<0.3 mg m〜(-3)。芥子气和路易斯矿1以及血药氰化氢和氯化氰被认为是“水疱”,其LOD分别为<2.4 mg m〜(-3)和> 415 mg m〜(-3)。氯蒸气被认为是“血”,LOD为820 mg m。氮芥末3和氯仿的蒸气根据其浓度被识别为不同的警报类别。氮芥1、2和光气的蒸气未显示任何警报。至于干扰,在检测的38种溶剂中,神经气体模拟物,二甲基甲基膦酸酯,三甲基磷酸酯,三乙基磷酸酯,二异丙基氟磷酸酯,起泡剂模拟物,2-氯乙基乙基硫醚,1,4-噻吨,2-巯基乙醇和20种有机溶剂的蒸气被认为是错误的。 -肯定。比较了6个离子迁移单元和半导体单元的传感器通道响应值的模式,并与化学战剂报警情况进行了比较。%化学剤(化学兵器用轨化学战剂)は。即rn效性で,微量で杀伤性,无能力性を示し,神経ガス,びrnらん剤,窒息剤,血液剤,くしゃみ剤,催涙剤などに分类rnされが。

著录项

  • 来源
    《分析化学》 |2010年第1期|65-76|共12页
  • 作者单位

    科学警察研究所法科学第三部:277-0882 千葉県柏市村の葉6-3-1;

    警視庁:100-8929 東京都千代田区霞ヶ関2-1-1;

    帝国繊維株式会社:103-0027 東京都中央区日本橋2-5-13;

    志賀有限会社: 227-0038 神奈川県横浜市青葉区奈良5-5-4;

    科学警察研究所法科学第三部:277-0882 千葉県柏市村の葉6-3-1;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 jpn
  • 中图分类
  • 关键词

相似文献

  • 外文文献
  • 中文文献