Hemostatic effects of Microporous Polysaccharide Hemosphere? in a rat model with severe femoral artery bleeding

机译:微孔多糖血球的止血作用?大鼠股动脉严重出血模型

获取原文
获取原文并翻译 | 示例
           

摘要

An appropriate hemostatic dressing for prehospital use should lower mortality due to uncontrolled hemorrhage. In this study, the investigators explored the hemostatic effects of Microporous Polysaccharide Hemosphere? (MPH) applied in a rat model with severe femoral artery bleeding. Twelve rats were randomly assigned to MPH and control groups: The femoral artery of each rat was pierced to initiate bleeding. Then, 0.25 g MPH was poured into the bleeding site. A 200-g scale weight was placed over the bleeding site for 30 sec. At 30-sec intervals, the scale weight was removed, and hemostasis was assessed visually. After 30 sec, if the bleeding had ceased, the test was scored and checked as “passed at 30 sec.” If the bleeding had not stopped, the same procedures were repeated a maximum of 3 times. If hemostasis could not be achieved even after the third application, the test was scored as failed. The same sequence of procedures was repeated for the control group without use of MPH and with only standard compression. Application of MPH resulted in complete control of bleeding in 2 of 6, 4 of 6, and 6 of 6 rats at 30, 60, and 90 sec, respectively. In the control group, however, hemostasis could not be achieved in all 6 rats, even at 90 sec. The difference between the 2 groups was statistically significant (P=.007). Application of MPH and compression with a scale weight significantly decreased the time of hemostasis in the rat model with femoral arterial bleeding.
机译:院前使用适当的止血敷料可降低失血导致的死亡率。在这项研究中,研究人员探讨了微孔多糖血球的止血作用。 (MPH)用于严重股动脉出血的大鼠模型。将12只大鼠随机分为MPH组和对照组:将每只大鼠的股动脉穿刺以开始出血。然后,将0.25g MPH倒入出血部位。将200克体重的重物放置在出血部位上30秒。每隔30秒,去除体重秤,并通过视觉评估止血效果。 30秒后,如果出血已停止,则对测试进行评分并检查为“ 30秒通过”。如果出血没有停止,则重复相同的步骤,最多3次。如果即使在第三次应用后仍无法止血,则将测试记为失败。对照组不使用MPH,仅进行标准压迫重复相同的程序顺序。 MPH的使用分别在30、60和90秒时完全控制了6只大鼠中的6只,4只,6只和6只的出血。然而,在对照组中,即使在90秒时,所有6只大鼠也未能实现止血。两组之间的差异具有统计学意义(P = .007)。 MPH的应用和体重秤的施压可显着减少股动脉出血大鼠模型的止血时间。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号