首页> 美国卫生研究院文献>Springer Open Choice >Influence of dose reduction and iterative reconstruction on CT calcium scores: a multi-manufacturer dynamic phantom study
【2h】

Influence of dose reduction and iterative reconstruction on CT calcium scores: a multi-manufacturer dynamic phantom study

机译:剂量减少和迭代重建对CT钙评分的影响:多制造商动态幻像研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

To evaluate the influence of dose reduction in combination with iterative reconstruction (IR) on coronary calcium scores (CCS) in a dynamic phantom on state-of-the-art CT systems from different manufacturers. Calcified inserts in an anthropomorphic chest phantom were translated at 20 mm/s corresponding to heart rates between 60 and 75 bpm. The inserts were scanned five times with routinely used CCS protocols at reference dose and 40 and 80% dose reduction on four high-end CT systems. Filtered back projection (FBP) and increasing levels of IR were applied. Noise levels were determined. CCS, quantified as Agatston and mass scores, were compared to physical mass and scores at FBP reference dose. For the reference dose in combination with FBP, noise level variation between CT systems was less than 18%. Decreasing dose almost always resulted in increased CCS, while at increased levels of IR, CCS decreased again. The influence of IR on CCS was smaller than the influence of dose reduction. At reference dose, physical mass was underestimated 3–30%. All CT systems showed similar CCS at 40% dose reduction in combinations with specific reconstructions. For some CT systems CCS was not affected at 80% dose reduction, in combination with IR. This multivendor study showed that radiation dose reductions of 40% did not influence CCS in a dynamic phantom using state-of-the-art CT systems in combination with specific reconstruction settings. Dose reduction resulted in increased noise and consequently increased CCS, whereas increased IR resulted in decreased CCS.
机译:要评估剂量减少与迭代重建(IR)组合对动态体模中不同制造商提供的最新CT系统的冠状动脉钙分数(CCS)的影响。拟人化的人体模型中的钙化插入物以20mm / s的速度翻译,对应于60至75bpm的心率。使用常规使用的CCS方案以参考剂量对插入物进行了五次扫描,并在四个高端CT系统上将剂量减少了40%和80%。应用了滤波反投影(FBP)和不断提高的IR水平。确定噪声水平。将CCS(以Agatston和质量得分量化)与物理质量和FBP参考剂量下的得分进行比较。对于结合FBP的参考剂量,CT系统之间的噪声水平变化小于18%。减少剂量几乎总是导致CCS升高,而在IR水平升高时,CCS再次降低。 IR对CCS的影响小于剂量减少的影响。在参考剂量下,身体质量被低估了3–30%。所有的CT系统在减少剂量40%的情况下与特定的重建相结合均显示出相似的CCS。对于某些CT系统,与IR结合使用时,CCS在剂量减少80%时不受影响。这项多供应商研究表明,使用最新的CT系统结合特定的重建设置,减少40%的辐射剂量不会影响动态体模中的CCS。剂量减少导致噪声增加,因此CCS增加,而IR增加导致CCS减少。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号