Volumetric measurement of dentoalveolar defects by means of intraoral 3D scanner and gravimetric model

机译:通过口内3D扫描仪和重量分析模型对牙槽骨缺损进行体积测量

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The aim of this study was to evaluate the accuracy in volumetric measurements obtained on an experimental model using an intraoral scanner and a gravimetric method. Three identical partial dentate maxillary acrylic models with three fabricated alveolar defects, in anterior and posterior regions, were scanned using an intraoral scanner (20 scans/defects). The defects differed in terms of size and distance of neighbouring teeth. As references, replicas of each defect were created using a dimensional stable silicone impression material. After measuring the mass of each replica, the volume was calculated by dividing the mass of each replica by the density of the impression material. The defects had a volume, according to the gravimetric method, ranging from 40.5 to 143.7 mm3. The scans were imported to metrology software for analyses. Accuracy was determined in terms of trueness and precision. The mean trueness for all defect types was 0.168 mm3 (SD 0.691, range 2.82). There was no statistical significant difference between the mean trueness for all defects measured (p = 0.910). The mean precision for all defect types was 0.147 mm3 (SD 0.524, range 2.86). There were no statistical significant differences between the dental models in regard to mean precision (p = 0.401), however, there were statistical significant differences between defects in position 1 and 2 (p = 0.002) and 1 and 3 (p = 0.001). Based on the findings of this study, the intraoral scanner utilized in the current study presented an acceptable level of accuracy when measuring volume of defects.
机译:这项研究的目的是评估使用口内扫描仪和重量分析法在实验模型上获得的体积测量的准确性。使用口腔内扫描仪扫描三个相同的部分齿状上颌丙烯酸模型,在前,后区域具有三个人造的肺泡缺损(20次扫描/缺陷)。缺陷在相邻牙齿的大小和距离方面有所不同。作为参考,使用尺寸稳定的有机硅印模材料创建了每个缺陷的副本。在测量每个复制品的质量之后,通过将每个复制品的质量除以压印材料的密度来计算体积。根据重量法,缺陷的体积为40.5至143.7mm 3 。扫描结果被导入到计量软件中进行分析。准确性是根据真实性和准确性确定的。所有缺陷类型的平均真实度为0.168mm 3 (SD 0.691,范围2.82)。所有测得缺陷的平均真实度之间没有统计学显着差异(p = 0.910)。所有缺陷类型的平均精度为0.147mm 3 (SD 0.524,范围2.86)。牙齿模型之间在平均精度方面没有统计学显着差异(p = 0.401),但是,位置1和2(p defects = 0.002)和1和3(p = 0.001)的缺陷之间也没有统计学上的显着差异。根据这项研究的结果,当前研究中使用的口腔内扫描仪在测量缺陷量时显示出可接受的准确性水平。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号