The purpose of this study was to apply texture analysis (TA) to evaluate the uniformity of SPECT images reconstructed with the 3D Ordered Subsets Expectation Maximization (3D-OSEM) algorithm. For this purpose, a cylindrical homogeneous phantom filled with 177Lu was used and a total of 24 spherical volumes of interest (VOIs) were considered inside the phantom. The location of the VOIs was chosen in order to define two different configurations, i.e. gravity and radial configuration. The former configuration was used to investigate the uniformity of distribution of 177Lu inside the phantom, while the latter configuration was used to investigate the lack of uniformity from center towards edge of the images. For each VOI, the trend of different texture features considered as a function of 3D-OSEM updates was investigated in order to evaluate the influence of reconstruction parameters. TA was performed using CGITA software. The equality of the average texture feature trends in both spatial configurations was assumed as the null hypothesis and was tested by functional analysis of variance (fANOVA). With regard to the gravity configuration, no texture feature rejected the null hypothesis when the number of subsets increased. For the radial configuration, the statistical analysis revealed that, depending on the 3D-OSEM parameters used, a few texture features were capable of detecting the non-uniformity of 177Lu distribution inside the phantom moving from the center of the image towards its edge. Finally, cross-correlation coefficients were calculated to better identify the features that could play an important role in assessing quality assurance procedures performed on SPECT systems.
机译:这项研究的目的是应用纹理分析(TA)评估使用3D有序子集期望最大化(3D-OSEM)算法重建的SPECT图像的均匀性。为此,使用了填充有 177 Lu的圆柱形均质体模,并且在体模内考虑了总共24个感兴趣的球形体积(VOI)。选择VOI的位置是为了定义两种不同的构型,即重力和径向构型。前者用于研究幻影内 177 Lu分布的均匀性,而后者用于研究从图像中心到边缘的均匀性不足。对于每种VOI,研究了作为3D-OSEM更新的函数的不同纹理特征的趋势,以评估重建参数的影响。使用CGITA软件执行TA。两种空间配置中平均纹理特征趋势的均等性均被视为原假设,并通过方差泛函分析(fANOVA)进行了测试。关于重力配置,当子集数量增加时,没有纹理特征会拒绝零假设。对于径向配置,统计分析表明,根据所使用的3D-OSEM参数,一些纹理特征能够检测幻影从中心移动的 177 Lu分布的不均匀性图像的边缘。最后,计算互相关系数以更好地识别在评估SPECT系统上执行的质量保证程序中可能起重要作用的特征。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号