首页> 美国卫生研究院文献>PLoS Clinical Trials >A novel gel liner system with embedded electrodes for use with upper limb myoelectric prostheses
【2h】

A novel gel liner system with embedded electrodes for use with upper limb myoelectric prostheses

机译:一种带有嵌入式电极的新型凝胶衬垫系统,用于上肢肌电假体

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

We present the development and evaluation of a gel liner system for upper limb prosthesis users that enables acquisition of electromyographic (myoelectric) control signals through embedded electrodes and flexible, conductive fabric leads. This liner system is constructed using a manufacturing approach rather than by modifying a commercially available liner. To evaluate the efficacy, eight male individuals with transhumeral amputations used this system, with standard myoelectric prostheses, for home trials lasting an average of 7.3 weeks. Before and after the home trials, electrical resistance of the cumulative 218 embedded electrodes and leads within 10 gel liner systems was measured and found to increase slightly (from an average of 13.4 to 27.5 Ω) after usage. While this increase was statistically significant (p = 0.001), all but one of the final resistance values remained low enough to enable consistent myoelectric control. User impressions were evaluated through a questionnaire comparing the liner prototypes to their own myoelectric prosthesis socket interface. Subjects preferred the liner prototype (p = 0.008) over their own system in the clinical areas of comfort, suspension, function, and, especially, ease of use. These results suggest that this gel liner system is a clinically viable option and that it may offer advantages over current clinical technology for users of upper limb myoelectric prostheses.
机译:我们介绍了针对上肢假肢用户的凝胶衬里系统的开发和评估,该系统能够通过嵌入式电极和柔性导电织物导线获取肌电(肌电)控制信号。使用制造方法而不是通过修改市售衬里来构造该衬里系统。为了评估疗效,八名经肱骨截肢的男性患者使用该系统和标准肌电假体进行家庭试验,平均持续7.3周。在家庭试验之前和之后,测量了10个凝胶衬垫系统中累积的218个嵌入式电极和引线的电阻,发现使用后电阻略有增加(从平均13.4到27.5Ω)。尽管这种增加在统计学上是显着的(p = 0.001),但除一个最终电阻值外,其他所有电阻值都保持足够低,可以进行稳定的肌电控制。通过将衬管原型与他们自己的肌电假体插槽接口进行比较的问卷调查,评估了用户的印象。在舒适性,悬吊性,功能性,尤其是易用性的临床方面,受试者偏爱衬套原型(p = 0.008)优于他们自己的系统。这些结果表明,该凝胶衬垫系统是临床上可行的选择,并且它可能为上肢肌电假体的使用者提供优于当前临床技术的优势。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号