Novel methods of testing and calibration of oscillometric blood pressure monitors

机译:示波血压计测试和校准的新方法

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

We present a robust method for testing and calibrating the performance of oscillometric non-invasive blood pressure (NIBP) monitors, using an industry standard NIBP simulator to determine the characteristic ratios used, and to explore differences between different devices. Assuming that classical auscultatory sphygmomanometry provides the best approximation to intra-arterial pressure, the results obtained from oscillometric measurements for a range of characteristic ratios are compared against a modified auscultatory method to determine an optimum characteristic ratio, Rs for systolic blood pressure (SBP), which was found to be 0.565. We demonstrate that whilst three Chinese manufactured NIBP monitors we tested used the conventional maximum amplitude algorithm (MAA) with characteristic ratios Rs = 0.4624±0.0303 (Mean±SD) and Rd = 0.6275±0.0222, another three devices manufactured in Germany and Japan either do not implement this standard protocol or used different characteristic ratios. Using a reference database of 304 records from 102 patients, containing both the Korotkoff sounds and the oscillometric waveforms, we showed that none of the devices tested used the optimal value of 0.565 for the characteristic ratio Rs, and as a result, three of the devices tested would underestimate systolic pressure by an average of 4.8mmHg, and three would overestimate the systolic pressure by an average of 6.2 mmHg.
机译:我们提供了一种用于测试和校准示波非侵入式血压(NIBP)监视器性能的可靠方法,使用行业标准的NIBP模拟器来确定所使用的特征比率,并探索不同设备之间的差异。假设经典听诊血压计能提供最佳的动脉内压近似值,则将示波法测量的一系列特征比率的结果与改良的听诊方法进行比较,以确定最佳的特征比率,即收缩压(SBP)的Rs,被发现是0.565。我们证明了,尽管我们测试的三台中国制造的NIBP监护仪使用常规的最大振幅算法(MAA),其特征比率Rs = 0.4624±0.0303(Mean±SD),Rd = 0.6275±0.0222,但另外三台在德国和日本制造的设备不执行此标准协议或使用不同的特征比率。使用来自102位患者的304条记录的参考数据库,其中包含Korotkoff声音和示波波形,我们显示测试的设备均未使用特征比Rs的最佳值0.565,因此,其中三个设备如果进行测试,收缩压平均会低估4.8mmHg,三个会高估收缩压平均6.2mmHg。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号