Percutaneous MR-guided interventions using an optical Moiré Phase tracking system: Initial results

机译:使用光学莫尔相位跟踪系统的经皮MR引导干预:初步结果

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The aim of this study was the development and evaluation of a real-time guidance support using optical Moiré Phase Tracking (MPT) for magnetic resonance (MR) guided percutaneous interventions. A gradient echo sequence, capable of real-time position updates by the MPT system, was modified to enable needle guidance based on four rigidly attached MPT markers at the back of a needle. Two perpendicular imaging planes were automatically aligned along the calibrated needle and centered at its tip. For user guidance, additional information about the needle trajectory and the tip to target distance were added as image overlay. Both, images and guiding information were displayed on the in-room monitor to facilitate MR guided interventions. The guidance support was evaluated by four experienced interventional radiologists and four novices targeting rubber O-rings embedded in a custom-made phantom on a 3T wide-bore MRI system (80 punctures). The skin to target time, user error, system error and total error were analyzed. The mean skin to target time was 146s±68s with no statistically significant difference between experts and novices. A low mean user error (0.91mm±0.43mm), system error (0.53mm±0.27mm) and total error (0.99mm±0.47mm) was reached in all directions. No statistically significant difference in user error, system error and total error could be found between experts and novices. The presented tracking and image guidance system combined with the user interface offers continuous and interactive control of the imaging plane while puncturing in the magnet enabling accurate real-time feedback for both, experienced and non-experienced users.
机译:这项研究的目的是开发和评估使用光学莫尔相位跟踪(MPT)进行磁共振(MR)引导的经皮干预的实时指导支持。可以通过MPT系统实时更新位置的梯度回波序列经过修改,可以基于针后部的四个刚性连接的MPT标记进行针引导。两个垂直的成像平面沿着校准的针自动对齐,并在其尖端居中。为了便于用户指导,添加了有关针迹轨迹和针尖到目标距离的其他信息作为图像叠加。图像和指导信息都显示在室内监视器上,以方便MR指导的干预。指导支持由四名经验丰富的介入放射科医生和四名新手针对在3T宽口MRI系统(80针)上定制模体中嵌入的橡胶O形圈进行评估。分析了目标时间的皮肤,用户错误,系统错误和总错误。皮肤到目标的平均时间为146s±68s,专家和新手之间在统计学上没有显着差异。在各个方向上均达到了较低的平均用户误差(0.91mm±0.43mm),系统误差(0.53mm±0.27mm)和总误差(0.99mm±0.47mm)。在专家和新手之间,在用户错误,系统错误和总错误方面没有统计上的显着差异。所提供的跟踪和图像引导系统与用户界面相结合,可对成像平面进行连续和交互式控制,同时刺入磁铁,可为有经验和无经验的用户提供准确的实时反馈。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号