首页> 美国卫生研究院文献>International Journal of Molecular Sciences >Continuous Flow Atmospheric Pressure Laser Desorption/Ionization Using a 6–7-µm-Band Mid-Infrared Tunable Laser for Biomolecular Mass Spectrometry
【2h】

Continuous Flow Atmospheric Pressure Laser Desorption/Ionization Using a 6–7-µm-Band Mid-Infrared Tunable Laser for Biomolecular Mass Spectrometry

机译:使用6-7μm波段中红外可调谐激光器对生物分子质谱进行连续流大气压激光解吸/电离

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A continuous flow atmospheric pressure laser desorption/ionization technique using a porous stainless steel probe and a 6–7-µm-band mid-infrared tunable laser was developed. This ion source is capable of direct ionization from a continuous flow with a high temporal stability. The 6–7-µm wavelength region corresponds to the characteristic absorption bands of various molecular vibration modes, including O–H, C=O, CH3 and C–N bonds. Consequently, many organic compounds and solvents, including water, have characteristic absorption peaks in this region. This ion source requires no additional matrix, and utilizes water or acetonitrile as the solvent matrix at several absorption peak wavelengths (6.05 and 7.27 µm, respectively). The distribution of multiply-charged peptide ions is extremely sensitive to the temperature of the heated capillary, which is the inlet of the mass spectrometer. This ionization technique has potential for the interface of liquid chromatography/mass spectrometry (LC/MS).
机译:开发了一种使用多孔不锈钢探头和6-7μm波段中红外可调谐激光器的连续流大气压激光解吸/电离技术。该离子源能够以高的时间稳定性从连续的流中直接电离。 6–7 µm波长区域对应于各种分子振动模式的特征吸收带,包括O–H,C = O,CH3和C–N键。因此,许多有机化合物和溶剂(包括水)在该区域具有特征性吸收峰。该离子源不需要额外的基质,并在几个吸收峰波长(分别为6.05和7.27 µm)处利用水或乙腈作为溶剂基质。带多个电荷的肽离子的分布对质谱仪入口即加热毛细管的温度极为敏感。这种电离技术对于液相色谱/质谱法(LC / MS)的界面具有潜力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号