Comparison of range of motion after standard and high-flexion posterior stabilised total knee replacement

机译:比较标准和高屈曲后路稳定全膝关节置换术后的运动范围

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The purpose of this study was to compare the range of motion after standard version posterior stabilised TKR and high-flexion version TKR in patients receiving bilateral total knee replacement. Thirty-five patients were recruited. The range of motion of the knees was measured clinically with a goniometer in both the pre-operative period and the most recent follow-up. It was found that the pre-operative range of motion was comparable in the two groups. The average post-operative flexion was 105° ± 13° in the standard version group and 106° ± 14° in the high-flexion design group (p = 0.201, paired t-test; beta error = 0.073). A slight loss in flexion was observed in the standard version group (0.5°) as opposed to a slight gain in the high-flexion design group (2°). However, this was not statistically significant (p = 0.251, paired t-test; beta error = 0.105).
机译:这项研究的目的是比较接受双侧全膝关节置换术的标准版本后稳定TKR和高屈曲版本TKR后的运动范围。招募了35名患者。在术前和最近的随访中,均用测角计对膝关节的运动范围进行了临床测量。发现两组的术前运动范围相当。标准版组的平均术后屈曲为105°±±13°,高屈曲设计组的平均术后屈曲为106°±±14°(p = 0.201,配对t检验;β误差= 0.073)。在标准版组(0.5°)中观察到轻微的屈曲损失,而在高屈曲设计组(2°)中观察到轻微的屈曲。然而,这在统计学上不显着(p = 0.251,配对t检验;β误差= 0.105)。