Selection of Internal Standards for Quantitative Matrix-AssistedLaser Desorption/Ionization Mass Spectrometric Analysis Based on CorrelationCoefficients

机译:定量基质辅助内标的选择基于相关性的激光解吸/电离质谱分析系数

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Matrix-assisted laser desorption/ionization mass spectrometry (MALDI MS) has shown its great success in the qualitative analysis of a wide range of organic and biological molecules. However, its application in quantitative analysis is still limited by the difficulty in the availability of isotope-labeled internal standards. The present work investigates the relationship between the correlation coefficient of the peak intensities of analyte and candidate internal standard ions and the linearity of possible quantitative analysis. Based on the two analyte examples, ciprofloxacin and substance P, the results show that the performance of the selected nonisotope-labeled internal standard is greatly related to the correlation coefficient. A high positive correlation coefficient (>0.7) between the ions of analyte and candidate standard can result in a good linearity (R2 > 0.98) and vice versa. The results provide a new way to select nonisotope-labeled internal standards for MALDI analysis and thus can be potentially applied in the rapid quantitative massspectrometry.
机译:基质辅助激光解吸/电离质谱(MALDI MS)在对多种有机和生物分子进行定性分析方面显示出巨大的成功。但是,其在定量分析中的应用仍然受到同位素标记内标物可用性的限制。本工作研究了分析物的峰强度与候选内标离子的相关系数与可能的定量分析的线性之间的关系。基于环丙沙星和P物质这两个分析物实例,结果表明,所选择的非同位素标记内标物的性能与相关系数密切相关。分析物离子与候选标准品之间的正相关系数高(> 0.7)时,线性关系较好(R 2

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号