PRELOAD AND TORQUE REMOVAL EVALUATION OF THREE DIFFERENT ABUTMENT SCREWS FOR SINGLE STANDING IMPLANT RESTORATIONS

机译:单点种植体修复的三种不同锚固螺钉的预紧力和扭矩去除评估

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Several authors still consider the mechanical problems of fracture and component loosening as the main causes of failure of implant-supported restorations. The purpose of this in vitro study was to compare the preload of three types of screw for transmucosal abutment attachment used in single implant-supported prosthesis through strain gauge and removal torque measurements. Three external hex fixtures were used, and each received a transmucosal abutment (Cera One®), which was fixed to the implant with its respective screw: Group A- gold screw, Group B- titanium screw and Group C- surface-treated titanium screw (Ti-Tite®). Ten screws of each type were attached applying a 30.07±0.28 Ncm torque force and maintained in position for 5 minutes. After this, the preload values were measured using strain gauges and a measurement cell. Gold screws presented higher preload values (131.72±8.98 N), followed by surface-treated titanium screws (97.78±4.68 N) and titanium screws (37.03±5.69 N). ANOVA (p<0.05) and Tukey's test (p<0.05) were applied. Statistically significant differences were found among the groups for both preload and removal torque values. In conclusion, gold screws may be indicated to achieve superior longevity of the abutment-implant connection and, consequently, prosthetic restoration due to greater preload values yielded.
机译:几位作者仍认为骨折的机械问题和部件松动是植入物支持的修复体失败的主要原因。这项体外研究的目的是通过应变仪和去除扭矩测量,比较三种类型的用于经粘膜基台附着的螺钉在单个植入物支持的假体中的预紧力。使用了三个外部六角固定装置,每个固定装置都接受了经粘膜基台(CeraOne®),并通过其各自的螺钉将其固定到植入物上:A组-金螺钉,B组-钛螺钉和C组-表面处理的钛螺钉(Ti-Tite®)。施加30.07±0.28 Ncm的扭力,连接10种每种类型的螺钉,并保持在原位5分钟。此后,使用应变仪和测量单元测量预载值。金螺丝的预紧力值更高(131.72±8.98 N),其次是表面处理的钛螺丝(97.78±4.68 N)和钛螺丝(37.03±5.69 N)。采用方差分析(p <0.05)和Tukey检验(p <0.05)。两组之间在预紧力和拆卸扭矩值上都有统计学上的显着差异。总而言之,可能需要使用金螺钉来实现基台与种植体连接的超长寿命,并且由于产生更大的预紧力值,因此可以修复假体。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号