On the evaluation of mobile target trajectory between four‐dimensional computer tomography and four‐dimensional cone‐beam computer tomography

机译:四维计算机断层扫描与四维锥形电脑断层扫描的移动目标轨迹的评价

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

For mobile lung tumors, four‐dimensional computer tomography (4D CT) is often used for simulation and treatment planning. Localization accuracy remains a challenge in lung stereotactic body radiation therapy (SBRT) treatments. An attractive image guidance method to increase localization accuracy is 4D cone‐beam CT (CBCT) as it allows for visualization of tumor motion with reduced motion artifacts. However, acquisition and reconstruction of 4D CBCT differ from that of 4D CT. This study evaluates the discrepancies between the reconstructed motion of 4D CBCT and 4D CT imaging over a wide range of sine target motion parameters and patient waveforms.
机译:对于移动肺肿瘤,四维计算机断层扫描(4D CT)通常用于仿真和治疗规划。本地化准确性仍然是肺部立体定向体放射治疗(SBRT)治疗的挑战。提高定位精度的有吸引力的图像引导方法是4D锥形光束CT(CBCT),因为它允许通过减少的运动伪影可视化肿瘤运动。但是,4D CBCT的获取和重建与4D CT的收购和重建不同。本研究评估了在各种正弦目标运动参数和患者波形上的4D CBCT和4D CT成像之间的重建运动与患者波形之间的差异。