首页> 美国卫生研究院文献>Annals of Vascular Diseases >Rupture of Abdominal Aortic Aneurysm Caused by Combined Type IIIb and Type Ia Endoleak with the Endurant II Endograft: A Case Report
【2h】

Rupture of Abdominal Aortic Aneurysm Caused by Combined Type IIIb and Type Ia Endoleak with the Endurant II Endograft: A Case Report

机译:组合IIIB型和IA型EndoLeak型腹主动脉瘤的破裂与跨厄的II endooRaft引起的:案例报告

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

We report a case of combined types IIIb and Ia endoleak that developed 6 years after endovascular aneurysm repair (EVAR) with the Endurant II® endograft for abdominal aortic aneurysm (AAA). The patient presented with post-EVAR AAA rupture and underwent emergency open repair. We observed types IIIb and Ia endoleak and successfully performed felt banding to preserve the stent graft. Type IIIb endoleak with the Endurant® endograft is rare, and treatments have not been fully established. We summarized the case reports regarding type IIIb endoleak with the Endurant® endograft and mainly discussed the treatments.
机译:我们举报了组合类型IIIB和IA Endoleak的案例,其在血管内动脉瘤修复(EVAR)后6年,腹主动脉瘤(AAA)的EndurantII®内传络。患者介绍了evar AAA破裂和接受紧急开放修复。我们观察到IIIB和IA Endoleak类型,并成功地进行了毡条以保持支架移植物。 IIIB型EndoLeak与Endurant®内传中稀有,并且尚未完全建立治疗方法。我们总结了关于IIIB型EndoLeak的案例报告与Endurant®EndooRoft,主要讨论了治疗方法。