Effects of a new shape‐memory alloy interspinous process device on pressure distribution of the intervertebral disc and zygapophyseal joints in vitro

机译:新型形状记忆合金棘突间处理装置对椎间盘和突关节压力分布的影响

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

>Objective:  To quantify the pressure distribution of lumbar intervertebral discs and zygapophyseal joints with different degrees of distraction of the interspinous processes by using a new shape‐memory interspinous process stabilization device, and to research the relationship between changing disc and zygapophyseal joint loads and the degree of distraction of interspinous processes, and thus optimize usage of the implant. >Methods:  Six cadaver lumbar specimens (L2–L5) were loaded. The loads in disc and zygapophyseal joints were recorded at each L3‐4 disc level. Implants with different spacer heights were then placed by turn and the pressure measurements repeated. >Results:  An implant with 10 mm spacer height does not significantly share the load. A 12 mm implant reduces the posterior annulus load, and meanwhile decreases the zygapophyseal joints pressure, but only in extension. A 14 mm implant shares the loads of posterior annulus, nucleus, and zygapophyseal joints in extension and the neutral position, but slightly increases the anterior annulus' load. Though 16–20 mm implants do decrease the loads in the posterior annulus and zygapophyseal joints, the anterior annulus' load was apparently increased. >Conclusion:  Different degrees of distraction of the interspinous processes lead to different load distribution on the intervertebral disc. The implant tested is not appropriate in cases of serious spinal stenosis because of the contradiction that, while over‐distraction of the interspinous processes decreases the posterior annulus and the zygapophyseal joints load and distracts the intervertebral foramina, it leads to a marked increase in the load of the anterior annulus, which is recognized to accelerate disc degeneration.
机译:>目的:通过使用新型的形状记忆型棘突间稳定装置,量化具有不同牵张程度的棘突间腰椎间盘和and突关节的压力分布,并研究换盘之间的关系关节突的负荷和棘突间的分散程度,从而优化植入物的使用。 >方法:装入六个尸体腰椎标本(L2-L5)。在每个L3-4椎间盘水平上记录椎间盘和突关节的负荷。然后依次放置具有不同垫片高度的植入物,并重复进行压力测量。 >结果::间隔高度为10mm的植入物不会明显分担负载。一个12mm的植入物可减轻后瓣环负荷,同时降低the突关节的压力,但仅限于伸展。一个14mm的植入物在伸展和中立位置分担后环,核和,突关节的负荷,但会稍微增加前环的负荷。尽管16–20mm的植入物确实减轻了后瓣环和突关节的负荷,但前瓣环的负荷显然增加了。 >结论:棘突间分散程度的不同会导致椎间盘的负荷分布不同。由于存在以下矛盾,所测试的植入物不适用于严重的椎管狭窄:尽管棘突间过度撑开会减少后环和the突关节负荷并分散椎间孔,但会导致负荷显着增加被认为会加速椎间盘退变。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号