Objective: The aim of this study was to evaluate whether stapes prostheses can be visualized with less metal artifacts and therefore more accurately on cone beam computed tomography in comparison to computed tomography imaging. Recent studies have shown that cone beam computed tomography has advantages when imaging metal artifacts. Patients with hearing loss or vertigo, who have undergone stapedotomy, often present a challenge for otologic surgeons. Imaging studies can deliver crucial additional information. Methods: A retrospective analysis of imaging studies and clinical data in a tertiary care center were carried out. Forty-one patients with forty-five implanted ears were evaluated in the study. All included patients had been implanted with a platinum–fluoroplastic (n = 19) or titanium (n = 26) piston and subsequently had undergone imaging months or years after surgery for various reasons. Patients underwent computed tomography or cone beam computed tomography of the temporal bone depending on availability. Piston visualization, prosthesis length, vestibular intrusion and audiologic results were compared between the groups. Piston position on imaging studies were compared to intraoperative findings. Results: Functional length measurements of all prostheses were carried out with a mean error of −0.17 mm (±0.20). Platinum–fluoroplastic protheses were significantly underestimated in length compared to titanium prostheses. To analyze the material-dependent difference in the measurement errors of the imaging techniques the interaction was tested in an ANOVA model and showed no statistically significant result (p = 0.24). The blinded neuroradiologist viewed two implants, both platinum–fluoroplastic pistons, as located outside of the vestibule due to an underestimation of the prothesis length and the missing radiodensity of the lower end of the prosthesis. Conclusion: Surgeons and radiologists should be aware of the different types and radiologic features of stapes prostheses and the missing radiodensity of some protheses parts. Cone beam computed tomography is an imaging alternative with a potential advantage of reduced radiation in patients after stapes surgery suffering from vertigo or hearing loss to evaluate piston position.
机译:目的:本研究的目的是评估芯片假体是否可以用较少的金属伪像可视化,因此与计算断层摄影成像相比,锥梁梁计算断层扫描。最近的研究表明,锥形光束计算机断层扫描在成像金属伪影时具有优势。患有听力丧失或眩晕的患者,他们经历过稳定的术,往往对耳科外科医生产生挑战。成像研究可以提供至关重要的其他信息。方法:对第三级护理中心的成像研究和临床数据进行回顾性分析。在研究中评估了四十五名植入耳的患者。所有包括患者均已植入铂 - 氟塑料(n = 19)或钛(n = 26)活塞,随后由于各种原因而经历了手术后的成像月份或数年。根据可用性,患者接受过计算的断层扫描或锥形束计算颞骨的断层扫描。在组之间比较了活塞可视化,假体长度,前庭侵扰和听力学结果。与术中发现进行了对成像研究的活塞位置。结果:所有假体的功能长度测量均由平均误差为0.17毫米(±0.20)。与钛假体相比,铂 - 氟塑料前导粒度明显低估。为了分析成像技术的测量误差的材料依赖性差异,在ANOVA模型中测试相互作用,并没有显示出统计学上显着的结果(P = 0.24)。由于低估了假体的下端,所揭示的神经皮层,铂 - 氟塑料活塞,两种植入物,两种植入物,铂 - 氟塑料活塞,如在前庭的外部,并且缺失假体的下端的缺失的射线。结论:外科医生和放射科医师应了解镫骨假体的不同类型和放射性特征,以及一些前列零件的缺失的射线。锥梁计算断层扫描是一种成像替代方案,其潜在的优势在患有眩晕或听力损失的患者中,患者减少患者辐射的潜在优势,以评估活塞位置。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号