Surface and build‐up dose comparison between Elekta 6 MV flattening filter and flattening‐filter‐free beams using an advanced Markus ionization chamber and a solid water‐equivalent phantom

机译:使用先进的马克电离室和固体水当量模拟的ELEKTA 6 MV扁平滤波器和无静滤梁的表面和积聚剂量比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Using a plane‐parallel advanced Markus ionization chamber and a stack of water‐equivalent solid phantom blocks, percentage surface and build‐up doses of Elekta 6 MV flattening filter (FF) and flattening‐filter‐free (FFF) beams were measured as a function of the phantom depth for field sizes ranging from 2 × 2 to 10 × 10 cm2. It was found that the dose difference between the FF and the FFF beams was relatively small. The maximum dose difference between the FF and the FFF beams was 4.4% at a depth of 1 mm for a field size of 2 × 2 cm2. The dose difference was gradually decreased while the field size was increased up to 10 × 10 cm2. The measured data were also compared to published Varian FF and FFF data, suggesting that the percentage surface and build‐up doses as well as the percentage dose difference between FF and FFF beams by our Elekta linac were smaller than those by the Varian linac.
机译:使用平面平行的先进马克电离室和一堆水当量的固体模块,测量Elekta 6 MV扁平滤波器(FF)和无扁平滤网(FFF)梁的百分比和堆积剂量用于场尺寸的幻像深度的功能范围从2×2到10×10 cm2。发现FF和FFF光束之间的剂量差相对较小。 FF和FFF梁之间的最大剂量差为4.4%,深度为1mm,用于2×2cm 2的场尺寸。剂量差异逐渐降低,而场尺寸增加至10×10cm 2。测量的数据也与发表的Varian FF和FFF数据进行比较,表明我们的ELEKTA LINAC的FF和FFF梁之间的百分比和积聚剂量以及FF和FFF光束之间的百分比差异小于Varian Linac的百分比。

著录项

相似文献

  • 外文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号