As immediate loading protocols are becoming more frequent, the primary stability of implants has become an essential criterion for the osseointegration of dental implants. Based on this, the objective of this study was to understand the influence of different undersized surgical preparation sites on the insertion torque (IT) and implant stability quotient (ISQ). Four different site-preparation protocols were performed on fresh humid type III bovine bone: one control, the standard protocol recommended by the manufacturer (P1), and three variations of undersized techniques (P2, P3 and P4). The implant used was VEGA by Klockner Implant System. The sample size was n = 40 for each of the four groups. A torquemeter was used to measure the IT, and the ISQ was measured with a Penguin RFA. Both variables showed a tendency to increase as the preparation technique was reduced, although not all the values were statistically significant (p < 0.05) when comparing with the standard preparation. The preparations without a cortical drill, P2 and P4, showed better results than those with a cortical drill. Given the limitations of this study, it can be concluded that reducing the implant preparation can increase both the IT and ISQ. Removing the cortical drill is an effective method for increasing implant stability, although it should be used carefully.
机译:由于立即加载方案变得更频繁,植入物的主要稳定性已成为牙科植入物骨整合的基本标准。基于这一点,本研究的目的是了解不同口径外科制剂位点对插入扭矩(IT)和植入物稳定价(ISQ)的影响。在新鲜潮湿型III牛骨上进行四种不同的位点制剂方案:一种控制,制造商推荐的标准协议(P1),以及虚线技术的三种变化(P2,P3和P4)。使用的植入物是Klockner植入系统的VEGA。四组中的每一个样本大小为n = 40。使用Tormemeter测量它,用企鹅RFA测量ISQ。随着制备技术降低,两种变量显示出增加,尽管与标准制备相比,但所有值并非所有值都是统计学意义的(P <0.05)。没有皮质钻头,P2和P4的制剂显示出比具有皮质钻头的钻头的结果更好。鉴于本研究的局限性,可以得出结论,减少植入物制剂可以增加它和ISQ。拆下皮质钻是一种有效的方法,用于提高植入物稳定性,尽管它应该仔细使用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号