An In Vitro Evaluation of Primary Stability Values for Two Differently Designed Implants to Suit Immediate Loading in Very Soft Bone

机译:两个不同设计的植入物的初级稳定性值的体外评估以适应非常柔软的骨骼的立即载荷

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。
获取外文期刊封面目录资料

摘要

The achievement of sufficient implant stability in poor quality bone seems to be a challenge. Most manufacturers develop special dental implants, which are claimed to show higher stability even in very soft bone. The aim of this experimental study was to compare two recently introduced dental implants with differing thread designs. A total of 11 implants of each group were inserted in the part of the fresh bovine ribs, corresponding to very soft bone. The primary stability was measured with resonance frequency analysis (RFA) and Periotest; the average of two measurements for each method and for each implant was taken and statistical analysis was applied. The highest stability values were obtained with the ICX Active Master implants, followed by the Conelog® Progressive-Line implants placed with the very soft bone protocol. The primary stability values of the Conelog® Progressive-Line implants inserted by the very soft bone protocol and the ICX Active Master implants placed with the standard protocol showed sufficient stability for immediate loading in low-density bone. Within the limitations of this study, the thread design of the implants and underdimensioned implant bed preparation seem to be effective for better primary stability in cancellous bone.
机译:在质量劣质骨骼中取得足够的植入稳定性似乎是一项挑战。大多数制造商开发了特殊的牙科植入物,这些牙科植入物即使在非常软骨中也表现出更高的稳定性。该实验研究的目的是比较最近引入的牙科植入物,具有不同的线程设计。将每组11种植入物插入新鲜牛肋的一部分中,对应于非常柔软的骨骼。用共振频率分析(RFA)测量初级稳定性和炎性;考虑了每种方法和每种植入物的两种测量的平均值,并施加统计分析。用ICX活性主植入物获得最高的稳定性值,然后用非常软骨协议放置Conelog®渐进线植入物。由非常软骨方案插入的Conelog®渐进式植入物的主要稳定性值和置于标准方案的ICX活性母植入物显示出足够的稳定性,用于在低密度骨中立即加载。在本研究的局限内,植入物的螺纹设计和底层植入床准备似乎有效地在松质骨中更好的初级稳定性有效。

著录项

相似文献

  • 外文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号