Cardiac troponin I is a marker for diagnosis of myocardial damage. Several immunoassays are currently available for determination of concentrations of troponin I in serum. We evaluated a chemiluminescent assay for troponin I using ACS:180 automated analyzer (Bayer Diagnostics). We compared our results with two other immunoassays using the OPUS Magnum (OPUS troponin I assay, Dade Behring) and AxSYM (microparticle enzyme immunoassay, Abbott) analyzers. The within‐run and between‐run CVs were less than 5% for all three levels of controls. The chemiluminescent assay for troponin I was linear up to a serum troponin I concentration of 50 ng/mL and the detection limit was 0.1 ng/mL of troponin. A good correlation between troponin I concentration measured by the chemiluminescent assay ( axis) and the microparticle enzyme immunoassay (MEIA) ( axis) was observed, although the concentrations of troponin I in individual specimens were approximately four times higher, when measured by the MEIA assay, than those measured by chemiluminescent assay. The correlation coefficient was 0.98 with the regression equation y = 0.22x + 1.125. We also observed a good correlation in troponin I concentrations obtained by the chemiluminescent assay ( axis) and OPUS troponin I assay ( axis). The correlation coefficient was 0.96 and the regression equation was y = 0.79x – 0.52. The correlation coefficient was 0.93 when we compared troponin I concentrations obtained by the OPUS assay ( axis) with the corresponding concentrations obtained by the MEIA assay ( axis). The corresponding regression equation was y = 0.25x + 3.5. We conclude that the chemiluminescent troponin I assay showed good analytical performance. J. Clin. Lab. Anal. 14:224–229, 2000. © 2000 Wiley‐Liss, Inc.
机译:心肌肌钙蛋白I是诊断心肌损伤的标志物。目前有几种免疫测定可用于测定血清中肌钙蛋白I的浓度。我们使用ACS:180自动分析仪(Bayer Diagnostics)评估了肌钙蛋白I的化学发光分析。我们将我们的结果与使用OPUS Magnum(OPUS肌钙蛋白I测定,Dade Behring)和AxSYM(微粒酶免疫测定,Abbott)分析仪的其他两种免疫测定进行了比较。对于所有三个级别的对照,运行内和运行间CV均小于5%。肌钙蛋白I的化学发光测定呈线性,直至血清肌钙蛋白I浓度为50 ng / mL,检出限为0.1 ng / mL肌钙蛋白。通过化学发光测定(轴)测得的肌钙蛋白I浓度与微粒酶免疫测定(MEIA)(轴)之间具有良好的相关性,尽管通过MEIA测定测得单个样品中的肌钙蛋白I浓度大约高四倍。 ,而不是化学发光法测得的结果。相关系数为0.98,回归方程y = 0.22x + 1.125。我们还观察到通过化学发光测定(轴)和OPUS肌钙蛋白I测定(轴)获得的肌钙蛋白I浓度具有良好的相关性。相关系数为0.96,回归方程为y = 0.79x – 0.52。当我们比较通过OPUS分析获得的肌钙蛋白I浓度(轴)与通过MEIA分析获得的相应浓度(轴)时,相关系数为0.93。相应的回归方程为y = 0.25x + 3.5。我们得出结论,化学发光肌钙蛋白I测定显示出良好的分析性能。 J.临床实验室肛门14:224–229,2000。©2000 Wiley-Liss,Inc.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号