Evaluation of noninvasive hemoglobin monitoring for hematological disorders

机译:无创血红蛋白监测血液学疾病的评估

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The Astrim is a noninvasive hemoglobin monitoring device that uses near‐infrared radiation. We studied the usefulness of this device for patients with hematological disorders. We measured hemoglobin levels 309 times in 140 subjects, using an automated hematology analyzer (K‐4500) and the Astrim. The coefficient of correlation between Hb levels with the K‐4500 (K‐Hb) and the Astrim (A‐Hb) was r=0.591 ( <0.001). The coefficient of correlation between A‐Hb and K‐Hb from 174 specimens with anemia and 135 without anemia was r=0.531 ( <0.001) and r=0.345 ( <0.001), respectively. In the 309 specimens, the Astrim showed a diagnostic sensitivity and specificity for anemia of 78.3% and 69.0%, respectively. While the r‐values, sensitivity, and specificity were passable, they appeared to be insufficient for an accurate evaluation. We believe this was caused by the measurement conditions (i.e., the finger selected for measurement, the finger position in relation to the CCD camera, and the finger temperature). On the other hand, the precision test results were good. Therefore, if careful attention is paid to measurement conditions, the noninvasive Astrim is clinically useful for continuous hemoglobin monitoring of hematological disorders and patients with hemorrhagic diseases, such as perioperative hemorrhage and gastrointestinal bleeding. However, it is difficult to determine Hb concentrations in one measurement using the Astrim because the r‐values, sensitivity, and specificity are insufficient for accurate evaluations of in‐patients, particularly patients with advanced diseases. Therefore, we believe it is necessary to improve the system so that A‐Hb can be accurately determined in one measurement. J. Clin. Lab. Anal. 19:1–5, 2005. © 2005 Wiley‐Liss, Inc.
机译:Astrim是一种使用近红外辐射的非侵入性血红蛋白监测设备。我们研究了该设备对血液系统疾病患者的有效性。我们使用自动血液分析仪(K-4500)和Astrim对140名受试者的血红蛋白水平进行了309次测量。 Hb水平与K-4500(K-Hb)和Astrim(A-Hb)之间的相关系数为r = 0.591(<0.001)。 174个有贫血的样本和135个无贫血的样本的A-Hb和K-Hb之间的相关系数分别为r = 0.531(<0.001)和r = 0.345(<0.001)。在309个标本中,Astrim对贫血的诊断敏感性和特异性分别为78.3%和69.0%。尽管r值,敏感性和特异性是可以通过的,但它们似乎不足以进行准确的评估。我们认为这是由测量条件(即选择进行测量的手指,相对于CCD摄像机的手指位置以及手指温度)引起的。另一方面,精度测试结果良好。因此,如果认真注意测量条件,则无创Astrim在临床上可用于连续监测血红蛋白,以监测血液系统疾病和出血性疾病,例如围手术期出血和胃肠道出血。但是,很难通过Astrim一次测定Hb浓度,因为r值,敏感性和特异性不足以准确评估住院患者,尤其是患有晚期疾病的患者。因此,我们认为有必要改进系统,以便一次测量即可准确确定A-Hb。 J.临床实验室肛门2005年9月1日至5日。©2005 Wiley-Liss,Inc.。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号