首页> 美国卫生研究院文献>Sensors (Basel Switzerland) >The Rehapiano—Detecting Measuring and Analyzing Action Tremor Using Strain Gauges
【2h】

The Rehapiano—Detecting Measuring and Analyzing Action Tremor Using Strain Gauges

机译:Rehapiano-使用应变片检测测量和分析动作震颤

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

We have developed a device, the Rehapiano, for the fast and quantitative assessment of action tremor. It uses strain gauges to measure force exerted by individual fingers. This article verifies the device’s capability to measure and monitor the development of upper limb tremor. The Rehapiano uses a precision, 24-bit, analog-to-digital converter and an Arduino microcomputer to transfer raw data via a USB interface to a computer for processing, database storage, and evaluation. First, our experiments validated the device by measuring simulated tremors with known frequencies. Second, we created a measurement protocol, which we used to measure and compare healthy patients and patients with Parkinson’s disease. Finally, we evaluated the repeatability of a quantitative assessment. We verified our hypothesis that the Rehapiano is able to detect force changes, and our experimental results confirmed that our system is capable of measuring action tremor. The Rehapiano is also sensitive enough to enable the quantification of Parkinsonian tremors.
机译:我们开发了Rehapiano设备,用于快速定量地评估动作震颤。它使用应变仪测量各个手指施加的力。本文验证了该设备测量和监视上肢震颤发展的能力。 Rehapiano使用精密的24位模数转换器和Arduino微型计算机,通过USB接口将原始数据传输到计算机以进行处理,数据库存储和评估。首先,我们的实验通过测量已知频率的模拟震颤验证了该设备。其次,我们创建了一个测量协议,该协议用于测量和比较健康患者和帕金森氏病患者。最后,我们评估了定量评估的可重复性。我们验证了Rehapiano能够检测到力变化的假设,并且我们的实验结果证实了我们的系统能够测量动作震颤。 Rehapiano也足够灵敏,可以量化帕金森氏震颤。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号