Comparison of HIV-1 RNA Measurements Obtained by Using Plasma and Dried Blood Spots in the Automated Abbott Real-Time Viral Load Assay

机译:在自动雅培实时病毒载量测定中使用血浆和干血斑获得的HIV-1 RNA测量值的比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Dried blood spots (DBS) may be a promising alternative specimen type to plasma for measuring the viral load (VL) in HIV-infected individuals in resource-limited settings. However, characterization of assay performance using DBS is incomplete. In this prospective study, the VL was measured in parallel using plasma and DBS specimens collected at the same time from 157 HIV-1-infected individuals. DBS were prepared by dispensing 50 μl of blood onto filter paper cards and were stored desiccated at −20°C. Nucleic acid extraction from plasma and DBS was performed automatically using the Abbott m2000sp instrument, and the VL was measured using the RealTime HIV-1 VL assay, which has a lower limit of detection of 40 HIV RNA copies/ml. The correlation between plasma and DBS results was good (R = 0.91; P < 0.001). The mean difference in the VL (DBS minus plasma) was 0.35 log copies (standard deviation [SD], 0.47 log copies). A total of 40 (26%) paired specimens had a difference of >0.5 log copy, and in 12 (7.8%) it was >1 log copy. the VL from DBS was measurable in 95.7% of specimens with a plasma VL of >2.74 log copies (550 HIV RNA copies/ml). In summary, the VL can reliably be measured using DBS with the Abbott RealTime HIV-1 assay. The estimated lower limit of detection of this automated methodology on DBS is 550 copies/ml, a threshold that may be acceptable for periodic VL monitoring in patients on antiretroviral therapy in resource-limited settings, where early detection of virologic treatment failure is often problematic.
机译:干血斑(DBS)可能是一种有希望的替代血浆的标本类型,可用于在资源有限的环境中测量HIV感染者的病毒载量(VL)。但是,使用DBS进行测定性能的表征是不完整的。在这项前瞻性研究中,使用同时从157名HIV-1感染者身上采集的血浆和DBS标本并行测量VL。通过将50μl血液分配到滤纸卡上来制备DBS,并将其在-20°C下干燥保存。使用Abbott m2000sp仪器自动从血浆和DBS中提取核酸,并使用RealTime HIV-1 VL测定法测量VL,其检测下限为40 HIV RNA拷贝/毫升。血浆与DBS结果之间的相关性良好(R = 0.91; P <0.001)。 VL(DBS减去血浆)的平均差异为0.35 log拷贝(标准偏差[SD],0.47 log拷贝)。总共40个(26%)配对样本的差异大于0.5个日志副本,其中12个(7.8%)的差异大于1个日志副本。在95.7%的样本中可测得DBS的VL,血浆VL> 2.74 log拷贝(550 HIV RNA拷贝/ ml)。总之,可以使用DBS和Abbott RealTime HIV-1测定法可靠地测量VL。在DBS上这种自动化方法的检测下限估计为550拷贝/ ml,对于在资源有限的环境中进行抗逆转录病毒治疗的患者进行定期VL监测可能是可以接受的阈值,在这种情况下,病毒治疗失败的早期检测常常会出现问题。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号