首页> 美国卫生研究院文献>Journal of Clinical Microbiology >Assessment on Experimental Bacterial Biofilms and in Clinical Practice of the Efficacy of Sampling Solutions for Microbiological Testing of Endoscopes
【2h】

Assessment on Experimental Bacterial Biofilms and in Clinical Practice of the Efficacy of Sampling Solutions for Microbiological Testing of Endoscopes

机译:内窥镜微生物测试的实验细菌生物膜评估和采样溶液功效的临床实践

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Opinions differ on the value of microbiological testing of endoscopes, which varies according to the technique used. We compared the efficacy on bacterial biofilms of sampling solutions used for the surveillance of the contamination of endoscope channels. To compare efficacy, we used an experimental model of a 48-h Pseudomonas biofilm grown on endoscope internal tubing. Sampling of this experimental biofilm was performed with a Tween 80-lecithin-based solution, saline, and sterile water. We also performed a randomized prospective study during routine clinical practice in our hospital sampling randomly with two different solutions the endoscopes after reprocessing. Biofilm recovery expressed as a logarithmic ratio of bacteria recovered on bacteria initially present in biofilm was significantly more effective with the Tween 80-lecithin-based solution than with saline solution (P = 0.002) and sterile water (P = 0.002). There was no significant difference between saline and sterile water. In the randomized clinical study, the rates of endoscopes that were contaminated with the Tween 80-lecithin-based sampling solution and the saline were 8/25 and 1/25, respectively (P = 0.02), and the mean numbers of bacteria recovered were 281 and 19 CFU/100 ml (P = 0.001), respectively. In conclusion, the efficiency and therefore the value of the monitoring of endoscope reprocessing by microbiological cultures is dependent on the sampling solutions used. A sampling solution with a tensioactive action is more efficient than saline in detecting biofilm contamination of endoscopes.
机译:对于内窥镜微生物测试的价值存在不同意见,具体取决于所使用的技术。我们比较了用于监测内窥镜通道污染的采样溶液在细菌生物膜上的功效。为了比较疗效,我们使用了在内窥镜内部管道上生长的48小时假单胞菌生物膜的实验模型。用基于吐温80卵磷脂的溶液,盐水和无菌水对该实验生物膜进行采样。我们还在常规的临床实践过程中,在我们医院的常规采样过程中,用两种不同的内窥镜解决方案在后处理后随机抽取了前瞻性研究。用吐温80卵磷脂基溶液比用盐溶液(P = 0.002)和无菌水(P = 0.002)表示,生物膜回收率表示为最初对生物膜中存在的细菌回收的细菌的对数比率,其效果要好得多。盐水和无菌水之间没有显着差异。在随机临床研究中,被吐温80卵磷脂基采样溶液和盐水污染的内窥镜感染率分别为8/25和1/25(P = 0.02),平均细菌回收率为分别为281和19 CFU / 100 ml(P = 0.001)。总之,通过微生物培养物监测内窥镜后处理的效率和价值取决于所使用的采样溶液。具有张力作用的采样溶液在检测内窥镜生物膜污染方面比盐水更有效。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号