首页> 美国卫生研究院文献>Journal of Clinical Microbiology >Comparison of the Sofia and Veritor Direct Antigen Detection Assay Systems for Identification of Influenza Viruses from Patient Nasopharyngeal Specimens
【2h】

Comparison of the Sofia and Veritor Direct Antigen Detection Assay Systems for Identification of Influenza Viruses from Patient Nasopharyngeal Specimens

机译:索非亚和Veritor直接抗原检测分析系统的比较用于从患者鼻咽标本中鉴定流感病毒

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Influenza antigen detection assays (Sofia fluorescent immunoassay [FIA] and Veritor) yield objective results, which are potentially useful for point-of-care testing. The assays were evaluated with reverse transcriptase PCR (RT-PCR) using 411 nasopharyngeal swab specimens. Sensitivity and specificity values (percentages) of 79.0/99.0 and 64.0/99.4 for influenza A and 92.9/96.7 and 78.6/98.7 for influenza B were obtained for the Sofia and Veritor assays, respectively.
机译:流感抗原检测测定法(Sofia荧光免疫测定法[FIA]和Veritor)可产生客观的结果,可能对即时检验有用。使用411项鼻咽拭子标本,通过逆转录酶PCR(RT-PCR)对测定进行评估。对于Sofia和Veritor分析,对于甲型流感,灵敏度和特异性值(百分比)分别为79.0 / 99.0和64.0 / 99.4,对乙型流感分别为92.9 / 96.7和78.6 / 98.7。