Inter-unit variability of multi-leaf collimator parameters for IMRT and VMAT treatment planning: a multi-institutional survey

机译:用于IMRT和VMAT治疗计划的多叶准直仪参数的单位间变异性:多机构调查

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Modern treatment machines have shown small inter-unit variability regarding beam data. Recently, vendor-provided average beam data, such as the Representative Beam Data (RBD) of the TrueBeam (Varian Medical Systems, Palo Alto, CA, USA), has been used for modeling of the Eclipse (Varian Medical Systems) treatment planning system. However, RBD does not provide multi-leaf collimator (MLC) parameters, such as MLC leaf transmission factor (LTF) and dosimetric leaf gap (DLG). We performed a web-based multi-institutional survey to investigate these parameters as well as the measurement protocols and customization of the parameters for intensity-modulated radiotherapy (IMRT) and/or volumetric modulated radiotherapy (VMAT) commissioning. We collected 69 sets of linear accelerator (linac) data from 58 institutions. In order to measure MLC parameters, most institutions used farmer-type ionization chambers with a sensitive volume of 0.6 cm , water phantoms, source surface distance of 90 cm with 10 cm depth, and a vendor-provided plan. The LTF showed small inter-unit variabilities, although the DLG showed large variations. For optimization of the parameters for IMRT/VMAT calculations, DLG values were upwardly adjusted at many institutions, whereas the LTF values were modestly changed. We clarified that MLC parameters were measured under the same conditions at more than half of the facilities. Most institutions customized parameters in a similar manner for IMRT/VMAT. The median measured and customized values obtained in our study will be valuable to verify MLC installation accuracy and to shorten the iterative processes of finding the optimal values.
机译:现代治疗仪已显示出有关射束数据的较小的单元间变异性。最近,供应商提供的平均光束数据,例如TrueBeam(Varian Medical Systems,Palo Alto,CA,美国)的代表光束数据(RBD),已用于Eclipse(Varian Medical Systems)治疗计划系统的建模。但是,RBD不提供多叶准直器(MLC)参数,例如MLC叶片透射因子(LTF)和剂量叶间隙(DLG)。我们进行了基于网络的多机构调查,以调查这些参数以及强度调制放射疗法(IMRT)和/或体积调制放射疗法(VMAT)调试的测量协议和参数定制。我们从58个机构收集了69套线性加速器(直线加速器)数据。为了测量MLC参数,大多数机构使用了农户型电离室,其敏感体积为0.6厘米,水体模,源表面距离为90厘米,深度为10厘米,并提供了厂商提供的计划。尽管DLG表现出较大的变化,但LTF的单位间变化较小。为了优化IMRT / VMAT计算的参数,在许多机构中向上调整了DLG值,而对LTF值进行了适度的更改。我们澄清说,MLC参数是在相同条件下对一半以上的设施进行测量的。大多数机构以类似的方式为IMRT / VMAT自定义参数。在我们的研究中获得的中值测量值和定制值将对验证MLC安装精度和缩短寻找最佳值的迭代过程非常有价值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号