Accuracy of a Point-of-Care Hemoglobin A1c Assay

机译:即时护理血红蛋白A1c分析的准确性

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Point-of-care test (POCT) HbA1c assays provide rapidly available results for clinical decision-making. Accuracy and precision must be established. Venous blood samples from 300 patients were assayed for HbA1c by a laboratory technician (“laboratory assessment”) with the POCT Alere Afinion™ assay and a laboratory (Premier Affinity ) assay. POCT results from 402 patients’ fingerstick samples assayed by nine nontechnician staff (“clinical assessment”) were compared with the laboratory assay. The laboratory assessment showed tight correlation ( = .977, < .001) between the assays. Mean absolute and relative differences were 0.01 percentage points and 2.1%, respectively. CVs for the POCT and laboratory assays were <2% and <1%, respectively. The clinical assessment also showed a tight correlation between the assays ( = .978, < .001), with mean absolute and relative differences of 0.2 percentage points and 3.41%, respectively. CV for the POCT assay was <2%. The POCT performed acceptably compared to the laboratory assay under realistic clinical conditions.
机译:即时检验(POCT)HbA1c分析可为临床决策提供快速可用的结果。必须建立准确性和精确度。实验室技术人员(“实验室评估”)使用POCT Alere Afinion™测定法和实验室(Premier Affinity)测定法对300例患者的静脉血样本中的HbA1c进行了测定。将九名非技术人员对402名患者指尖样本进行的POCT结果(“临床评估”)与实验室分析进行了比较。实验室评估显示各测定之间存在紧密的相关性(= .977,<.001)。平均绝对和相对差异分别为0.01个百分点和2.1%。 POCT和实验室分析的CV分别为<2%和<1%。临床评估还显示,测定之间存在紧密的相关性(= .978,<.001),其绝对绝对值和相对差分别为0.2个百分点和3.41%。 POCT分析的CV <2%。与实际临床条件下的实验室分析相比,POCT的性能可以接受。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号