Defibrillation failure with short ICD charge time: Internal short circuit of a Durata lead

机译:ICD充电时间短的除颤失败:Durata导线内部短路

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A 17-year-old male with hypertrophic cardiomyopathy underwent placement of a dual-chamber implantable cardioverter defibrillator (ICD) in 2010. In October 2016, he suffered a cardiac arrest. His ICD was interrogated after visiting our hospital. The shock failed, and ventricular fibrillation (VF) terminated spontaneously to sinus rhythm. The Durata lead (SJM, Sylmar, CA, USA) was removed for investigation. The lead had an internal abrasion of a right ventricular conductor cable at the superior vena cava coil. These findings suggest that the ICD shocks that failed to terminate the VF were delivered with low energy due to an internal short circuit of the Durata lead. The very short charge time indicates insufficient energy delivery, which may cause failure of cardioversion and defibrillation.
机译:一名患有肥厚型心肌病的17岁男性在2010年接受了双室植入式心脏复律除颤器(ICD)的治疗。2016年10月,他因心脏骤停而心跳骤停。他的ICD在去我们医院后受到了讯问。休克失败,心室纤颤(VF)自发终止为窦性心律。杜拉塔铅(SJM,Sylmar,CA,USA)被移走进行调查。导线在上腔静脉线圈处的右心室导体电缆内部磨损。这些发现表明,由于杜拉塔引线的内部短路,未能终止VF的ICD冲击以低能量传递。极短的充电时间表示能量输送不足,可能导致心脏复律和除颤失败。