Is Post-Reprocessing Microbiological Surveillance of Duodenoscopes Effective in Reducing the Potential Risk in Transmitting Pathogens?

机译:十二指肠镜的后处理微生物监测是否能有效降低传播病原体的潜在风险?

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The use of a contaminated endoscope may lead to infections due to the transmission of potential pathogens from patient to patient. Post-reprocessing microbiological surveillance of four duodenoscopes was carried out over a three-year period in the Digestive Endoscopy Unit of an Italian hospital. Sampling of duodenoscopes was performed after the devices have been reprocessed. The initial phase of surveillance involved the contemporary evaluation of the four duodenoscopes; afterwards, microbiological surveillance proceeded at monthly intervals. The initial phase of surveillance revealed that three duodenoscopes presented a high level of contamination with “high-concern” micro-organisms, some of which were multi-drug-resistant. The highest values of contamination regarded the species (2500 CFU/duodenoscope), (2580 CFU/duodenoscope), and (2600 CFU/duodenoscope). Since the cultures were repeatedly positive on three successive occasions, the contaminated devices were sent to the manufacturer for evaluation. Audits were carried out with the personnel responsible for reprocessing, which was aimed to optimize the procedures used, and subsequently, only one case of non-conformity was found. Our study highlighted both the potential risk of transmitting pathogens through the use of duodenoscopes and the importance of implementing a well-structured system of microbiological surveillance and training programs, in order to reduce the risk of spreading retrograde cholangiopancreatography (ERCP)-associated infections.
机译:由于潜在病原体在患者之间的传播,使用受污染的内窥镜可能会导致感染。在意大利一家医院的消化内窥镜科进行了为期三年的对四个十二指肠镜的后处理微生物监测。在对设备进行重新处理后,对十二指肠镜进行采样。监视的初始阶段涉及对四个十二指肠镜的当代评估。之后,每月进行一次微生物监测。监视的初始阶段显示,三个十二指肠镜对“高度关注”的微生物造成了高水平的污染,其中有些具有多药耐药性。污染的最高值考虑的种类是(2500 CFU /十二指肠镜),(2580 CFU /十二指肠镜)和(2600 CFU /十二指肠镜)。由于培养物已连续三次重复呈阳性,因此将受污染的设备发送给制造商进行评估。对负责后处理的人员进行了审核,目的是优化所使用的程序,随后,仅发现了一个不合格的案例。我们的研究强调了通过使用十二指肠镜传播病原体的潜在风险,以及实施结构良好的微生物监测和培训计划系统的重要性,以降低传播逆行胰胆管造影(ERCP)相关感染的风险。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号