Agreement of Three Posturographic Force Plates in the Assessment of Postural Stability

机译:姿势稳定性评估中三个姿势性测力板的协议

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

This study was designed to assess how the results obtained for three different posturographic platforms agreed with each other in an assessment of static postural stability. The study included 111 young healthy participants. A measurement of postural stability was made for each participant, with their eyes open and then closed, on each platform in a random order. The Romberg ratio, path length, and center of pressure (COP) area were analyzed. For all measures, significant differences ( < 0.05) were observed among the three force plates. The highest Spearman’s rank correlation was observed between Alfa vs. CQStab2P (0.20 to 0.38), and the lowest between Alfa vs. AccuGait (−0.19 to 0.09). Similar results were obtained for the concordance correlation coefficient (0.10 to 0.22 for Alfa vs. CQStab2P and −0.6 to 0.02 for Alfa vs. AccuGait). Bland-Altman analysis for values standardized (z-scores) against AccuGait indicated a low level of agreement between compared platforms, with the largest error between AccuGait vs. Alfa, and a slightly lower error between AccuGait vs. CQStab2P or Alfa vs. CQStab2P. The 95% limits of agreement ranged from 2.38 to 7.16 (Alfa vs. AccuGait), 2.09 to 5.69 (CQStab2P vs. AccuGait), and 1.39 to 7.44 (AccuGait vs. Alfa) in COP length with eyes open and COP length Romberg ratio, respectively. Special care is recommended when comparing values relating to COPs from different devices that are analyzed by different software. Moreover, unperturbed stance tests among young healthy adults can be questioned as a valid postural control parameter.
机译:本研究旨在评估在评估静态姿势稳定性时,三种不同姿势平台获得的结果如何彼此一致。该研究包括111名年轻健康参与者。在每个平台上以随机顺序对每个参与者进行睁眼然后闭眼的姿势稳定性测量。分析了Romberg比率,路径长度和压力中心(COP)区域。对于所有测量,在三个测力板之间观察到显着差异(<0.05)。在Alfa与CQStab2P之间观察到的Spearman等级相关性最高(0.20至0.38),在Alfa与AccuGait之间的相关性最低(-0.19至0.09)。对于一致性相关系数也获得了类似的结果(Alfa与CQStab2P的对比值为0.10至0.22,Alfa与AccuGait的对比值为-0.6至0.02)。针对AccuGait标准化值(z分数)的Bland-Altman分析表明,比较平台之间的一致性较低,AccuGait与Alfa之间的误差最大,而AccuGait与CQStab2P或Alfa与CQStab2P的误差略低。睁大眼睛时的COP长度和COP长度Romberg比值的95%一致性限制范围为2.38至7.16(Alfa对AccuGait),2.09至5.69(CQStab2P对AccuGait)和1.39至7.44(AccuGait对Alfa)。分别。在比较与来自不同设备的COP的值(由不同软件分析)进行比较时,建议特别小心。此外,年轻健康成年人中不受干扰的姿势测试可被视为有效的姿势控制参数。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号