首页> 美国卫生研究院文献>Journal of Clinical Microbiology >Multicenter Evaluation of Fully Automated BACTEC Mycobacteria Growth Indicator Tube 960 System for Susceptibility Testing of Mycobacterium tuberculosis
【2h】

Multicenter Evaluation of Fully Automated BACTEC Mycobacteria Growth Indicator Tube 960 System for Susceptibility Testing of Mycobacterium tuberculosis

机译:全自动BACTEC结核分枝杆菌生长指示剂960系统对结核分枝杆菌药敏试验的多中心评估

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The reliability of the BACTEC Mycobacteria Growth Indicator Tube (MGIT) 960 system for testing of Mycobacterium tuberculosis susceptibility to the three front-line drugs (isoniazid [INH], rifampin [RIF], and ethambutol [EMB]) plus streptomycin (STR) was compared to that of the BACTEC 460 TB system. The proportion method was used to resolve discrepant results by an independent arbiter. One hundred and ten strains were tested with an overall agreement of 93.5%. Discrepant results were obtained for seven strains (6.4%) with INH (resistant by BACTEC MGIT 960; susceptible by BACTEC 460 TB), for one strain (0.9%) with RIF (resistant by BACTEC MGIT 960; susceptible by BACTEC 460 TB), for seven strains (6.4%) with EMB (six resistant by BACTEC MGIT 960 and susceptible by BACTEC 460 TB; one susceptible by BACTEC MGIT 960 and resistant by BACTEC 460 TB), and for 19 strains (17.3%) with STR (resistant by BACTEC MGIT 960 and susceptible by BACTEC 460 TB). After resolution of discrepant results, the sensitivity of the BACTEC MGIT 960 system was 100% for all four drugs and specificity ranged from 89.8% for STR to 100% for RIF. Turnaround times were 4.6 to 11.7 days (median, 6.5 days) for BACTEC MGIT 960 and 4.0 to 10.0 days (median, 7.0 days) for BACTEC 460 TB. These data demonstrate that the fully automated and nonradiometric BACTEC MGIT 960 system is an accurate method for rapid susceptibility testing of M. tuberculosis.
机译:用于测试结核分枝杆菌对三种一线药物(异烟肼[INH],利福平[RIF]和乙胺丁醇[EMB])和链霉素(STR)的敏感性的BACTEC分枝杆菌生长指示器管(MGIT)960系统的可靠性是与BACTEC 460 TB系统相比。比例法用于由独立仲裁员解决不一致的结果。测试了110个菌株,总体一致性为93.5%。对于具有INH的7株菌株(6.4%)(对BACTEC MGIT 960耐药;对BACTEC 460 TB敏感),对于对RIF的一种菌株(0.9%)(对BACTEC MGIT 960耐药;对BACTEC 460 TB敏感)获得了不同的结果,含EMB的7株(6.4%)(对BACTEC MGIT 960有抵抗力的6株,对BACTEC 460 TB易感;对BACTEC MGIT 960敏感并且对BACTEC 460 TB的耐药性的19株)(对BACTEC 460 TB有抗性) BACTEC MGIT 960,易受BACTEC 460 TB的影响)。解决了不一致的结果后,BACTEC MGIT 960系统对所有四种药物的敏感性均为100%,特异性范围从STR的89.8%到RIF的100%。 BACTEC MGIT 960的周转时间为4.6至11.7天(中位数为6.5天),BACTEC 460 TB的周转时间为4.0至10.0天(中位数为7.0天)。这些数据表明,全自动,非放射线的BACTEC MGIT 960系统是用于快速诊断结核分枝杆菌的准确方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号