An Analysis of Errors in Kinematic Parameters Associated with in Vivo Functional Radiographs

机译:体内功能X线片相关运动学参数的误差分析

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A pair of functional radiographs, taken at each end of the range of motion, are used to determine spinal motions. Graphic construction and computer-assisted methods are available for the radiographic analysis. The latter provides many more motion parameters. A study of lumbar spine lateral radiographs was conducted to determine errors in the motion parameters due to spinal level, radiographic quality, and errors in the two digitizing instruments. Significant differences were found in the errors due to the two digitizers when the same radiographic pair was redigitized several times. There were only minimal differences, however, between the digitizers when the radiographic films were remarked and redigitized. The error ranges (2 × SD) for the motion parameters were 1) rotation = ± 1.25°; 2) translation of the inferior posterior vertebral body corner = ± 0.86°; and 3) coordinates for the center of rotation = ± 4.3 mm. Both the spinal level and radiographic quality affected the magnitude of errors in all motion parameters.
机译:在运动范围的每一端拍摄的一对功能射线照片用于确定脊柱运动。图形构造和计算机辅助方法可用于射线照相分析。后者提供了更多的运动参数。进行了腰椎侧位X射线照片的研究,以确定由于脊柱水平,射线照相质量和两种数字化仪器中的误差引起的运动参数误差。当相同的射线照相对被重新数字化几次时,由于两个数字化仪而导致的错误中发现了显着差异。但是,当对胶片进行标记和重新数字化时,数字转换器之间只有极小的差异。运动参数的误差范围(2×SD)为:1)旋转=±1.25°; 2)椎体后下角的平移=±0.86°;和3)旋转中心的坐标=±4.3毫米。脊柱水平和射线照相质量都影响所有运动参数中误差的大小。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号