首页> 美国卫生研究院文献>Journal of Clinical Microbiology >Discrepancies between results by E-test and standard microbroth dilution testing of Streptococcus pneumoniae for susceptibility to vancomycin.
【2h】

Discrepancies between results by E-test and standard microbroth dilution testing of Streptococcus pneumoniae for susceptibility to vancomycin.

机译:肺炎链球菌对万古霉素的敏感性的E-检验结果与标准微肉汤稀释试验结果之间存在差异。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Vancomycin susceptibility testing of Streptococcus pneumoniae by the E-test consistently resulted in MICs that were at the upper limit of the 1-microgram/ml susceptible category defined by current National Committee for Clinical Laboratory Standards guidelines and were always higher than MICs obtained by microbroth dilution. Three of five E-test results for S. pneumoniae ATCC 49619 were higher than the acceptable limits.
机译:通过E检验对万古霉素对肺炎链球菌的药敏试验始终导致MIC达到当前国家临床实验室标准委员会指南所定义的1微克/毫升易感类别的上限,并且始终高于通过微孔稀释获得的MIC 。肺炎链球菌ATCC 49619的五项E检验结果中有三项高于可接受的限值。