Signal Enhancement in HPLC/Micro-Coil NMR Using Automated Column Trapping

机译:HPLC / Micro-Coil NMR中使用自动色谱柱捕获实现的信号增强

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A new HPLC-NMR system is described that performs analytical separation, pre-concentration, and NMR spectroscopy in rapid succession. The central component of our method is the online pre-concentration sequence that improves the match between post-column analyte peak volume and the micro-coil NMR detection volume. Separated samples are collected on to a C18 guard column with a mobile phase composed of 90% D2O/10% acetonitrile-D3, and back-flashed to the NMR micro-coil probe with 90% acetonitrile-D3/10% D2O. In order to assess the performance of our unit, we separated a standard mixture of 1 mM ibuprofen, naproxen, and phenylbutazone using a commercially available C18 analytical column. The S/N measurements from the NMR acquisitions indicated that we achieved signal enhancement factors up to 10.4 (±1.2)-fold. Furthermore, we observed that pre-concentration factors increased as the injected amount of analyte decreased. The highest concentration enrichment of 14.7 (±2.2)-fold was attained injecting 100 μL solution of 0.2 mM (~4 μg) ibuprofen.
机译:描述了一种新的HPLC-NMR系统,该系统可以快速连续进行分析分离,预浓缩和NMR光谱分析。我们方法的主要组成部分是在线预浓缩序列,可改善柱后分析物峰体积与微线圈NMR检测体积之间的匹配度。将分离的样品收集到带有由90%D2O / 10%乙腈-D3组成的流动相的C18保护柱上,并用90%乙腈-D3 / 10%D2O闪回NMR微线圈探针。为了评估我们装置的性能,我们使用市售的C18分析柱分离了1 mM布洛芬,萘普生和苯基丁酮的标准混合物。来自NMR采集的S / N测量表明,我们实现了高达10.4(±1.2)倍的信号增强因子。此外,我们观察到,随着分析物进样量的减少,预浓缩因子增加。注入0.2μM(〜4μg)布洛芬100μL溶液可获得最高14.7(±2.2)倍的浓度富集。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号