首页> 美国卫生研究院文献>other >A Novel Transcranial Magnetic Stimulator Inducing Near Rectangular Pulses with Controllable Pulse Width (cTMS)
【2h】

A Novel Transcranial Magnetic Stimulator Inducing Near Rectangular Pulses with Controllable Pulse Width (cTMS)

机译:一种新颖的经颅磁刺激器可诱导具有可控脉冲宽度(cTMS)的近矩形脉冲

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A novel transcranial magnetic stimulation (TMS) device with controllable pulse width (PW) and near rectangular pulse shape (cTMS) is described. The cTMS device uses an insulated gate bipolar transistor (IGBT) with appropriate snubbers to switch coil currents up to 7 kA, enabling PW control from 5 μs to over 100 μs. The near-rectangular induced electric field pulses use 22–34% less energy and generate 67–72% less coil heating compared to matched conventional cosine pulses. CTMS is used to stimulate rhesus monkey motor cortex in vivo with PWs of 20 to 100 μs, demonstrating the expected decrease of threshold pulse amplitude with increasing PW. The technological solutions used in the cTMS prototype can expand functionality, and reduce power consumption and coil heating in TMS, enhancing its research and therapeutic applications.
机译:描述了一种新型的经颅磁刺激(TMS)装置,其具有可控的脉冲宽度(PW)和接近矩形的脉冲形状(cTMS)。 cTMS器件使用带有适当缓冲器的绝缘栅双极晶体管(IGBT)来切换高达7 kA的线圈电流,从而使PW控制从5μs变为100μs以上。与匹配的传统余弦脉冲相比,近矩形感应电场脉冲使用的能量减少了22–34%,线圈产生的热量减少了67–72%。 CTMS用于在体内以20至100μs的PW刺激恒河猴的运动皮层,表明随着PW的增加,阈值脉冲幅度的预期下降。 cTMS原型中使用的技术解决方案可以扩展功能,并减少TMS中的功耗和线圈发热,从而增强其研究和治疗应用。