首页> 美国卫生研究院文献>other >Tension and fatigue behavior of 316LVM 1×7 multi-strand cables used as implantable electrodes
【2h】

Tension and fatigue behavior of 316LVM 1×7 multi-strand cables used as implantable electrodes

机译:用作植入电极的316LVM 1×7多股电缆的拉伸和疲劳行为

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The mechanical behavior of 316LVM 1×7 cables were evaluated in uniaxial tension, and in cyclic strain-controlled fatigue with the use of a Flex tester operated to provide fully reversed bending fatigue. The magnitude of cyclic strains imparted to each cable tested was controlled via the use of different diameter mandrels. Smaller diameter mandrels produced higher values of cyclic strain and lower fatigue life. Multiple samples were tested and analyzed via scanning electron microscopy. The fatigue results were analyzed via a Coffin-Manson-Basquin approach and compared to fatigue data obtained from the literature where testing was conducted on similar materials, but under rotating bending fatigue conditions.
机译:316LVM 1×7电缆的机械性能在单轴张力和循环应变控制疲劳中进行了评估,使用的Flex测试仪可提供完全反向的弯曲疲劳。通过使用不同直径的心轴,可以控制施加到每个测试电缆上的循环应变的大小。较小的心轴会产生较高的循环应变值,并降低疲劳寿命。通过扫描电子显微镜测试和分析多个样品。疲劳结果通过Coffin-Manson-Basquin方法进行了分析,并与从文献中获得的疲劳数据进行了比较,在类似的材料上进行了测试,但在旋转弯曲疲劳条件下进行了测试。