Assessing validity and reliability of Resting Metabolic Rate in six gas analysis systems

机译:评估六种气体分析系统中静止代谢率的有效性和可靠性

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The Deltatrac Metabolic Monitor (DTC), one of the most popular indirect calorimetry systems for measuring resting metabolic rate (RMR) in human subjects, is no longer being manufactured. This study compared five different gas analysis systems to the DTC. Resting metabolic rate was measured by the DTC and at least one other instrument at three study sites for a total of 38 participants. The five indirect calorimetry systems included: MedGraphics CPX Ultima, MedGem, Vmax Encore 29 System, TrueOne 2400, and Korr ReeVue. Validity was assessed using paired t-tests to compare means while reliability was assessed by using both paired t-tests and root mean square calculations with F tests for significance. Within-subject comparisons for validity of RMR revealed a significant difference between the DTC and Ultima. Bland-Altman plot analysis showed significant bias with increasing RMR values for the Korr and MedGem. Respiratory exchange ratio (RER) analysis showed a significant difference between the DTC and the Ultima and a trend for a difference with the Vmax (p = 0.09). Reliability assessment for RMR revealed that all instruments had a significantly larger coefficient of variation (CV) (ranging from 4.8% to 10.9%) for RMR compared to the 3.0 % CV for the DTC. Reliability assessment for RER data showed none of the instrument CV’s were significantly larger than the DTC CV. The results were quite disappointing, with none of the instruments equaling the within person reliability of the DTC. The TrueOne and Vmax were the most valid instruments in comparison with the DTC for both RMR and RER assessment. Further testing is needed to identify an instrument with the reliability and validity of the DTC.
机译:Deltatrac代谢监测仪(DTC)是用于测量人类受试者静息代谢率(RMR)的最流行的间接量热系统之一,现已不再生产。这项研究将五种不同的气体分析系统与DTC进行了比较。通过DTC和至少一种其他仪器在三个研究地点对总共38名参与者进行了静息代谢率的测量。五个间接量热系统包括:MedGraphics CPX Ultima,MedGem,Vmax Encore 29系统,TrueOne 2400和Korr ReeVue。有效性通过配对t检验比较均值,而可靠性通过配对t检验和均方根计算以及F检验评估显着性。 RMR有效性的受试者内部比较显示出DTC和Ultima之间存在显着差异。 Bland-Altman图分析显示,Korr和MedGem的RMR值增加,偏差显着。呼吸交换率(RER)分析显示DTC和Ultima之间存在显着差异,并且与Vmax有差异(p = 0.09)。 RMR的可靠性评估表明,与DTC的CV的3.0%CV相比,所有仪器的RMR的变异系数(CV)都很大(从4.8%到10.9%)。对RER数据的可靠性评估表明,任何仪器的CV都不比DTC CV大得多。结果非常令人失望,没有任何一种工具能够等同于DTC的内部可靠性。与DTC相比,TrueOne和Vmax是RMR和RER评估中最有效的工具。需要进一步测试以识别具有DTC可靠性和有效性的仪器。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号