The effect of repeated torque tightening on total lengths of implant abutments in different internal implant‒abutment connections

机译:反复拧紧扭矩对不同内部种植体-基台连接中种植体基台总长度的影响

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

>Background. Since the misfit of crown has an important role in clinical performance of implant-supported prostheses, and due to the impact of the settling effect on misfit, the aim of this study was to investigate the impact of torque forces on the total lengths of narrow and short implant abutments in different internal implant‒abutment connections. >Methods. In four different implant‒abutment connections, 8 analog implants with a normal diameter (4 mm) and narrow abutment (4.5 mm) were selected from groups of internal hex, internal octagon, morse hex 6° and morse hex 11°. Each of them was mounted within plaster type IV, and 32 samples were obtained. Then, the amount of vertical displacement was measured by closing the impression copings and applying torques of 20 25 and 30 Ncm. This stage was repeated for the abutment. In the next stage, the resin pattern was built and measurements were performed after applying the torques mentioned. Finally, after making the frame, this stage was repeated, and the settling effect was statistically analyzed with ANOVA. >Results. In the stages of impression coping, resin pattern and final prosthesis, HEXAGONE had significantly the highest and OCTAGONE had the lowest rates of settling, and the settling of morse hex 11° and 6° was between them. >Conclusion. Octagon implant had significantly the lowest settling in various clinical and laboratory stages by applying different torques.
机译:>背景。由于牙冠的错位在植入物支持的假体的临床表现中具有重要作用,并且由于沉降效应对错位的影响,因此本研究的目的是调查牙冠的影响。在不同的内部种植体-基台连接中,扭矩对狭窄和短种植体基台总长度的影响。 >方法。在四个不同的种植体-基台连接中,从内六角,内八角形,莫氏六角6°组中选择了8个正常直径(4 mm)和窄基台(4.5 mm)的模拟种植体。莫尔斯六角形为11°。将它们各自安装在IV型灰泥中,并获得32个样品。然后,通过关闭压印帽并施加20 25和30 Ncm的扭矩来测量垂直位移量。对基台重复此阶段。在下一阶段,建立树脂图案并在施加上述扭矩后进行测量。最后,制成框架后,重复此阶段,并使用ANOVA对沉降效果进行统计分析。 >结果。在压印应对,树脂图案和最终假体阶段,HEXAGONE的沉降率最高,而OCTAGONE的沉降率最低,莫氏六角形的11°和6°沉降率介于两者之间。 。 >结论。在不同的临床和实验室阶段,八角形植入物通过施加不同的扭矩具有最低的沉降。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号