首页> 美国卫生研究院文献>Journal of Diabetes Science and Technology >Evaluating the Performance of a Novel Embedded Closed-loop System
【2h】

Evaluating the Performance of a Novel Embedded Closed-loop System

机译:评估新型嵌入式闭环系统的性能

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The objective was to assess the reliability of a novel automated closed-loop glucose control system developed within the AP@home consortium in adults with type 1 diabetes. Eight adults with type 1 diabetes on insulin pump therapy (3 men; ages 40.5 ± 14.3 years; HbA1c 8.2 ± 0.8%) participated in an open-label, single-center, single-arm, 12-hour overnight study performed at the clinical research facility. A standardized evening meal (80 g CHO) accompanied by prandial insulin boluses were given at 19:00 followed by an optional snack of 15 g at 22:00 without insulin bolus. Automated closed-loop glucose control was started at 19:00 and continued until 07:00 the next day. Basal insulin delivery (Accu-Chek Spirit, Roche) was automatically adjusted by Cambridge model predictive control algorithm, running on a purpose-built embedded device, based on real-time continuous glucose monitor readings (Dexcom G4 Platinum). Closed-loop system was operational as intended over 99% of the time. Overnight plasma glucose levels (22:00 to 07:00) were within the target range (3.9 to 8.0 mmol/l) for 75.4% (37.5, 92.9) of the time without any time spent in hypoglycemia (<3.9 mmol/l). Mean overnight glucose was 7.8 ± 1.3 mmol/l. For the entire 12-hour closed-loop period (19:00 until 07:00) plasma glucose levels were within the target range (3.9 to 10.0 mmol/l) for 84.4% (63.3, 100) of time. There were no adverse events noted during the trial. We observed a high degree of reliability of the automated closed-loop system. The time spent in target glucose level overnight was comparable to results of previously published studies. Further developments to miniaturize the system for home studies are warranted.
机译:目的是评估在AP @ home联盟内部开发的新型1型糖尿病成人自动闭环葡萄糖控制系统的可靠性。八名接受胰岛素泵治疗的1型糖尿病成人(3名男性;年龄40.5±14.3岁; HbA1c 8.2±0.8%)参加了在临床上进行的开放性,单中心,单臂,12小时通宵研究研究设施。在19:00提供标准化的晚餐(80 g CHO),并伴以餐前胰岛素推注,然后在22:00任选15g零食,不添加胰岛素推注。自动化的闭环血糖控制始于19:00,一直持续到第二天的07:00。基础剑桥的胰岛素输送量(Accu-Chek Spirit,Roche)由Cambridge模型预测控制算法自动调整,并基于实时连续血糖监测仪读数(Dexcom G4 Platinum)在专用嵌入式设备上运行。闭环系统在99%的时间内都可以正常运行。过夜血糖水平(22:00至07:00)在目标范围(3.9至8.0 mmol / l)内的时间为75.4%(37.5,92.9),而无任何时间用于低血糖(<3.9 mmol / l) 。平均过夜葡萄糖为7.8±1.3 mmol / l。在整个12小时闭环时间段(19:00至07:00)中,血糖水平在目标范围内(3.9至10.0 mmol / l),持续时间为84.4%(63.3,100)。试验期间未发现不良事件。我们观察到了自动化闭环系统的高度可靠性。过夜达到目标血糖水平的时间与以前发表的研究结果相当。进一步发展以使家庭学习系统小型化是必要的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号