Volume assessment accuracy in computed tomography: a phantom study

机译:计算断层扫描中的批量评估准确性:幻影研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

There is a broad push in the cancer imaging community to eventually replace linear tumor measurements with three-dimensional evaluation of tumor volume. To evaluate the potential accuracy of volume measurement in tumors by CT, a gelatin phantom consisting of 55 polymethylmethacrylate (PMMA) spheres spanning diameters from 1.6 mm to 25.4 mm was fabricated and scanned using thin slice (0.625 mm) CT (GE LightSpeed 16). Nine different reconstruction combinations of field of view dimension (FOV = 20, 30, 40 cm) and CT kernel (standard, lung, bone) were analyzed. Contiguous thin-slice images were averaged to produce CT images with greater thicknesses (1.25, 2.50, 5.0 mm). Simple grayscale thresholding techniques were used to segment the PMMA spheres from the gelatin background, where a total of 1800 spherical volumes were evaluated across the permutations studied. The geometric simplicity of the phantom established upper limits on measurement accuracy. In general, smaller slice thickness and larger sphere diameters produced more accurate volume assessment than larger slice thickness and smaller sphere diameter. The measured volumes were smaller than the actual volumes by a common factor depending on slice thickness; overall, 0625 mm slices produced on average 18%, 1.25 mm slices produced 22%, 2.5 mm CT slices produced 29%, and 5.0 mm slices produced 39% underestimates of volume (mm3). Field of view did not have a significant effect on volume accuracy. Reconstruction algorithm significantly affected volume accuracy (p < 0.0001), with the lung kernel having the smallest error, followed by the bone and standard kernels. The results of this investigation provide guidance for CT protocol development and may guide the development of more advanced techniques to promote quantitatively accurate CT volumetric analysis of tumors.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号