The objective was to assess the effects of short-term (≤1 hour) static high temperature and humidity stresses on the performance of point-of-care (POC) glucose test strips and meters. Glucose meters are used by medical responders and patients in a variety of settings including hospitals, clinics, homes, and the field. Reagent test strips and instruments are potentially exposed to austere environmental conditions. Glucose test strips and meters were exposed to a mean relative humidity of 83.0% (SD = 8.0%) and temperature of 42°C (107.6°F, SD = 3.2) in a Tenney BTRC environmental chamber. Stressed and unstressed glucose reagent strips and meters were tested with spiked blood samples (n = 40 measurements per time point for each of 4 trials) after 15, 30, 45, and 60 minutes of exposure. Wilcoxon’s signed rank test was applied to compare measurements test strip and meter measurements to isolate and characterize the magnitude of meter versus test strip effects individually. Stressed POC meters and test strips produced elevated glucose results, with stressed meter bias as high as 20 mg/dL (17.7% error), and stressed test strip bias as high as 13 mg/dL (12.2% error). The aggregate stress effect on meter and test strips yielded a positive bias as high as 33 mg/dL (30.1% error) after 15 minutes of exposure. Short-term exposure (15 minutes) to high temperature and humidity can significantly affect the performance of POC glucose test strips and meters, with measurement biases that potentially affect clinical decision making and patient safety.
机译:目的是评估短期(≤1小时)静态高温和湿度应力对即时医疗点(POC)葡萄糖测试条和仪表性能的影响。医护人员和患者在各种场合(包括医院,诊所,家庭和野外)都使用血糖仪。试剂试纸和仪器可能会暴露于恶劣的环境条件下。在Tenney BTRC环境室内,将葡萄糖测试条和血糖仪的平均相对湿度为83.0%(SD = 8.0%),温度为42°C(107.6°F,SD = 3.2)。暴露15、30、45和60分钟后,用加标血样(n = 4次试验中每个时间点40次测量)测试受压和未受压的葡萄糖试剂条和测量仪。 Wilcoxon的带符号等级测试用于比较测试条和仪表的测量结果,以分别隔离和表征仪表对测试条效果的大小。压力POC仪和试纸产生的葡萄糖结果升高,压力仪偏倚高达20 mg / dL(17.7%误差),压力试纸偏倚高达13 mg / dL(12.2%误差)。暴露15分钟后,对仪表和试纸条的总应力影响产生高达33 mg / dL(30.1%误差)的正偏差。短期(15分钟)暴露于高温和高湿度会严重影响POC葡萄糖试纸和血糖仪的性能,其测量偏差可能会影响临床决策和患者安全。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号