Performance Variability of Seven Commonly Used Self-Monitoring of Blood Glucose Systems: Clinical Considerations for Patients and Providers

机译:七个常用的血糖系统自我监测的性能变异性:患者和提供者的临床注意事项

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

BackgroundBlood glucose data are frequently used in clinical decision making, thus it is critical that self-monitoring of blood glucose (SMBG) systems consistently provide accurate results. Concerns about SMBG accuracy have prompted the development of newly proposed International Organization for Standardization (ISO) standards: ≥95% of individual glucose results shall fall within ±15 mg/dl of the results of the manufacturer’s reference procedure at glucose concentrations <100 mg/dl and within ±15% for values ≥100 mg/dl. We evaluated seven marketed systems against the current and proposed ISO criteria (criterion A).
机译:背景技术血糖数据经常用于临床决策,因此至关重要的是,血糖自我监测(SMBG)系统始终如一地提供准确的结果。对SMBG准确性的担忧促使新提议的国际标准化组织(ISO)标准得以发展:在葡萄糖浓度<100 mg / g时,≥95%的单个葡萄糖结果应在制造商参考程序结果的±15 mg / dl范围内对于≥100mg / dl,应在dl和±15%以内。我们根据当前和提议的ISO标准(标准A)评估了七个上市系统。