The aim of this study was to complete a full evaluation of the new EIZO RX850 liquid crystal display and compare it to two currently used medical displays in Australia (EIZO GS510 and Barco MDCG 5121). The American Association of Physicists in Medicine (AAPM) Task Group 18 Quality Control test pattern was used to assess the performance of three high-resolution primary medical displays: EIZO RX850, EIZO GS510, and Barco MDCG 5121. A Konica Minolta spectroradiometer (CS-2000) was used to assess luminance response, non-uniformity, veiling glare, and color uniformity. Qualitative evaluation of noise was also performed. Seven breast lesions were displayed on each monitor and photographed with a calibrated 5.5-MP Olympus E-1 digital SLR camera. ImageJ software was used to sample pixel information from each lesion and surrounding background to calculate their conspicuity index on each of the displays. All monitor fulfilled all AAPM acceptance criteria. The performance characteristics for EIZO RX850, Barco MDCG 5121, and EIZO GS510 respectively were as follows: maximum luminance (490, 500.5, and 413 cd/m2), minimum luminance (0.724, 1.170, and 0.92 cd/m2), contrast ratio (675:1, 428:1, 449:1), just-noticeable difference index (635, 622, 609), non-uniformity (20, 5.92, and 8.5 %), veiling glare (GR = 2465.6, 720.4, 1249.8), and color uniformity (Δu′v′ = +0.003, +0.002, +0.002). All monitors demonstrated low noise levels. The conspicuity index (χ) of the lesions was slightly higher in the EIZO RX850 display. All medical displays fulfilled AAPM performance criteria, and performance characteristics of EIZO RX850 are equal to or better than those of the Barco MDCG 5121 and EIZO GS510 displays.
机译:这项研究的目的是完成对新型EIZO RX850液晶显示器的全面评估,并将其与澳大利亚目前使用的两种医疗显示器(EIZO GS510和Barco MDCG 5121)进行比较。美国医学物理学会(AAPM)任务组18质量控制测试图案用于评估三种高分辨率的主要医学显示器的性能:EIZO RX850,EIZO GS510和Barco MDCG5121。柯尼卡美能达光谱仪(CS- (2000年)用于评估亮度响应,不均匀,遮盖眩光和颜色均匀性。还对噪声进行了定性评估。每个监视器上显示七个乳腺病变,并用经过校准的5.5-MP Olympus E-1数字SLR相机拍照。使用ImageJ软件对每个病变和周围背景的像素信息进行采样,以计算每个显示器上的显着性指数。所有监控器均符合所有APM验收标准。 EIZO RX850,Barco MDCG 5121和EIZO GS510的性能特征分别为:最大亮度(490、500.5和413cd / m 2 ),最小亮度(0.724、1.170和0.92) cd / m 2 ),对比度(675:1、428:1、449:1),差异系数恰到好处(635、622、609),不均匀性(20、5.92,和8.5%),面纱眩光(GR = 2465.6、720.4、1249.8)和颜色均匀度(Δu'v'= +0.003,+0.002,+0.002)。所有显示器均显示出低噪音水平。在EIZO RX850显示器中,病变的显眼指数(χ)稍高。所有医疗显示器均符合AAPM性能标准,并且EIZO RX850的性能特征等于或优于Barco MDCG 5121和EIZO GS510显示器。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号