A pilot study of eye movement during mammography interpretation: Eyetracker results and workstation design implications

机译:乳房X光检查解释期间眼睛运动的初步研究:Eyetracker结果和工作站设计含义

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Digital mammography can potentially improve mammography image and interpretation quality. On-line interpretation from a workstation may improve interpretation logistics and increase availability of comparison images. Interpretation of eight 4k- × 5k-pixel mammograms on two to four 2k- × 2.5k-pixel monitors is problematic because of the time spent in choosing which images to display on which monitors, and zooming and roaming on individual images that are too large to display completely at full resolution. The authors used an eyetracker to measure radiologists viewing behavior during mammography interpretation with film on a viewbox. It was observed that a significant portion of the mammographers' time is spent viewing “comparison pairs” (typically two or more comparisons per case), such as the left mediolateral and craniocaudal images or old and new images. From the eyetracker measurements, we estimated that the number of images display, roam, and zoom operations decreases from an average of 64 for one monitor to 31 for four monitors, with the largest change going from one to two monitors. We also show that fewer monitors with a faster response time is superior to more monitors with a slower response time. Finally, the authors demonstrate the applicatity of time-motion analysis to mammographic workstation design.
机译:数字化乳腺X线照相术可以潜在地改善乳腺X线照相术的图像和解释质量。来自工作站的在线口译可以改善口译后勤工作,并提高比较图像的可用性。在两到四个2k-2.5k像素的监视器上解释八张4k-×5k像素的乳房X光照片是有问题的,因为在选择要在哪个监视器上显示哪些图像以及缩放太大的单个图像上花费时间以完整的分辨率完全显示。作者使用眼动仪来测量放射科医生在乳房X光检查期间使用视盒上的胶片进行解释的行为。据观察,乳房X光检查人员的大部分时间都花在查看“比较对”上(每个案例通常进行两次或更多次比较),例如左中外侧图像和颅尾图像或新旧图像。根据眼动仪的测量,我们估计图像显示,漫游和缩放操作的数量从一台显示器的平均数量64个减少到四台显示器的31个数量,最大变化是从一台显示器到两台显示器。我们还显示,响应时间较短的显示器要少于响应时间较慢的显示器。最后,作者演示了时间运动分析在乳腺X线摄影机设计中的应用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号