The purpose of this work is to correct for transient gradient waveform errors in magnetic resonance imaging (MRI), whether from eddy currents, group delay, or gradient amplifier nonlinearities, which are known to affect image quality. An iterative method is proposed to minimize error between desired and measured gradient waveforms, whose success does not depend on accurate knowledge of the gradient system impulse response. The method was applied to half-pulse excitation for 2-D ultra-short echo time (UTE) imaging on a small animal MRI system and to spiral 2-D excitation on a human 7T MRI system. Predistorted gradient waveforms reduced temporal signal variation caused by excitation gradient trajectory errors in 2-D UTE, and improved the quality of excitation patterns produced by spiral excitation pulses. Iterative gradient predistortion is useful for minimizing transient gradient errors without requiring accurate characterization of the gradient system impulse response.
机译:这项工作的目的是纠正磁共振成像(MRI)中的瞬态梯度波形误差,无论是由于涡流,群延迟还是梯度放大器非线性影响的,这些误差均会影响图像质量。提出了一种迭代方法来最小化所需和测量的梯度波形之间的误差,该方法的成功与否取决于梯度系统脉冲响应的准确知识。该方法适用于在小型动物MRI系统上进行2-D超短回波时间(UTE)成像的半脉冲激发,以及在人7T MRI系统上用于螺旋2-D激发。预失真的梯度波形减少了二维UTE中由激发梯度轨迹误差引起的时间信号变化,并改善了螺旋激发脉冲产生的激发模式的质量。迭代梯度预失真可用于最大程度地减小瞬态梯度误差,而无需精确表征梯度系统脉冲响应。