The goal of this study was to assess the effect of corneal hydration on the quality of the femtosecond laser (FSL) anterior lamellar cut. The Visumax FSL was used to dissect an 8-mm-diameter corneal flap in 22 eye bank corneas showing various levels of hydration. The intended ablation depth was 220 µm in all eyes, which corresponded to the maximal depth available with this laser. After the cut, the achieved ablation depth was measured using optical coherence tomography images, flap separability was assessed by measuring the mean force generated to detach the flap, and stromal bed roughness was assessed by measuring the Haralick contrast level on the 1000× scanning electron microscopy images of the ablated surfaces. The preoperative central corneal thickness ranged from 547 to 1104 µm (mean ± SEM: 833±30 µm). A negative correlation was found between the level of corneal hydration and the ablation depth measured in the mid-peripheral cornea (r = −0.626, p = 0.003), the ablation being more superficial in more edematous corneas. The Haralick contrast also tended to increase as a function of corneal hydration (r = 0.416, p = 0.061), suggesting that laser ablation in edematous corneas results in rougher stromal surfaces. These results support the hypothesis that the quality of the FSL lamellar cut decreases as the level of corneal hydration increases. Although FSL is still considered in the field as the tool of the future for corneal dissection, a better understanding of the limits of this tool will be needed before it can replace manual or automated stromal dissection techniques in hydrated corneas.
机译:这项研究的目的是评估角膜水化对飞秒激光(FSL)前薄片切割质量的影响。 Visumax FSL用于在22个眼库角膜中解剖直径8毫米的角膜瓣,显示出不同程度的水合作用。所有眼睛的预期消融深度均为220 µm,与该激光可获得的最大深度相对应。切开后,使用光学相干断层扫描图像测量获得的消融深度,通过测量产生的分离皮瓣的平均力来评估皮瓣的可分离性,并通过在1000倍扫描电子显微镜上测量Haralick对比水平来评估基质床的粗糙度烧蚀表面的图像。术前中央角膜厚度范围为547至1104 µm(平均±SEM:833±30 µm)。角膜水合程度与周围周围角膜的消融深度之间呈负相关(r = -0.626,p = 0.003),消融在水肿性角膜中更浅。 Haralick造影剂也倾向于随着角膜水化作用而增加(r = 0.416,p = 0.061),这表明在水肿角膜中进行激光消融会导致较粗糙的基质表面。这些结果支持以下假设:FSL片状切口的质量随着角膜水合水平的增加而降低。尽管FSL在该领域仍被认为是未来角膜剥离的工具,但在取代水合角膜中的手动或自动基质剥离技术之前,需要对该工具的局限性有更好的了解。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号