Observations Of Electrical Coupling Index Using the Contact™ System During Pulmonary Vein Electrical Isolation Procedures

机译:在肺静脉电隔离过程中使用Contact™系统观察电耦合指数

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The Contact (St Jude Medical) System uses a novel impedance- based measure of Electrical Coupling Index (ECI) to assess the quality of catheter tip to endocardium contact. We sought to establish average ECI measurements and behaviour during pulmonary vein (PV) isolation procedures.Forty-five patients undergoing PV isolation for atrial fibrillation (AF) were studied.‘Non-contact’ and upper range ‘in-contact’ catheter positioning was performed for system calibration. ECI measurements were recorded pre-ablation at 14 standardized locations around the PV antra.The mean ECI non-contact value was 77 ± 11 (range 63–107); the mean upper range in-contact value was 111 ± 16 (range 81–145). Mean ECI values pre-ablation around the PV antra ranged from 85 ± 18 to 107 ± 19. A trend towards higher mean ECI values was noted with increasing body mass index (BMI). Pre-ablation mean ECI values were 92 ± 10 (BMI 20-25), 95 ± 12 (BMI 26-30) and 104 ± 11 (BMI >30) (p< 0.01 for 20-25 vs. >30). A positive correlation was noted for mean pre-ablation ECI values and BMI (r=0.50).An expected range of ECI values during PV isolation has been documented in this study. Observed ECI values correlate with patient BMI. The potential limitations of the current generation Contact System and scope for future clinical applications are discussed.
机译:接触(St Jude Medical)系统使用基于阻抗的新颖电耦合指数(ECI)度量来评估导管尖端与心内膜接触的质量。我们试图建立隔离肺静脉(PV)过程中的平均ECI测量值和行为。研究了45位因房颤(AF)进行PV隔离的患者。执行系统校准。 ECI测量记录在消融前在PV角膜周围的14个标准化位置。平均ECI非接触值为77±11(范围63-107);平均ECI非接触值为77±11。平均范围上限接触值为111±16(范围81–145)。 PV消融前的平均消融前ECI值在85±18至107±19之间。随着体重指数(BMI)的增加,平均ECI值趋于较高。消融前的平均ECI值为92±10(BMI 20-25),95±12(BMI 26-30)和104±11(BMI> 30)(20-25与> 30的p <0.01)。消融前平均ECI值与BMI呈正相关(r = 0.50)。PV隔离期间ECI值的预期范围已在本研究中得到了证明。观察到的ECI值与患者的BMI相关。讨论了当前接触系统的潜在局限性以及未来临床应用的范围。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号