首页> 美国卫生研究院文献>other >Comparison of Peak Cardiopulmonary Performance Parameters on aRobotics-Assisted Tilt Table a Cycle and a Treadmill
【2h】

Comparison of Peak Cardiopulmonary Performance Parameters on aRobotics-Assisted Tilt Table a Cycle and a Treadmill

机译:心肺功能峰值参数的比较机器人辅助倾斜台脚踏车和跑步机

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Robotics-assisted tilt table (RATT) technology provides body support, cyclical stepping movement and physiological loading. This technology can potentially be used to facilitate the estimation of peak cardiopulmonary performance parameters in patients who have neurological or other problems that may preclude testing on a treadmill or cycle ergometer. The aim of the study was to compare the magnitude of peak cardiopulmonary performance parameters including peak oxygen uptake (VO2peak) and peak heart rate (HRpeak) obtained from a robotics-assisted tilt table (RATT), a cycle ergometer and a treadmill. The strength of correlations between the three devices, test-retest reliability and repeatability were also assessed. Eighteen healthy subjects performed six maximal exercise tests, with two tests on each of the three exercise modalities. Data from the second tests were used for the comparative and correlation analyses. For nine subjects, test-retest reliability and repeatability of VO2peak and HRpeak were assessed. Absolute VO2peak from the RATT, the cycle ergometer and the treadmill was (mean (SD)) 2.2 (0.56), 2.8 (0.80) and 3.2 (0.87) L/min, respectively (p < 0.001). HRpeak from the RATT, the cycle ergometer and thetreadmill was 168 (9.5), 179 (7.9) and 184 (6.9) beats/min, respectively (p< 0.001). VO2peak and HRpeak from the RATT vs thecycle ergometer and the RATT vs the treadmill showed strong correlations.Test-retest reliability and repeatability were high for VO2peak andHRpeak for all devices. The results demonstrate that the RATT isa valid and reliable device for exercise testing. There is potential for theRATT to be used in severely impaired subjects who cannot use the standardmodalities.
机译:机器人辅助倾斜台(RATT)技术提供身体支撑,周期性踏步运动和生理负荷。该技术可潜在地用于促进神经系统或其他问题的患者心肺功能峰值参数的估计,这些问题可能会妨碍在跑步机或自行车测功机上进行测试。这项研究的目的是比较峰值心肺功能参数的大小,包括从机器人辅助倾斜台(RATT),自行车测功机和跑步机获得的峰值摄氧量(VO2peak)和峰值心率(HRpeak)。还评估了三个设备之间的相关强度,重测可靠性和可重复性。 18名健康受试者进行了6次最大运动测试,其中对3种运动方式分别进行了2次测试。来自第二次测试的数据用于比较和相关性分析。对于9名受试者,评估了VO2peak和HRpeak的重测信度和重复性。来自RATT,自行车测功机和跑步机的绝对VO2峰值分别为(平均值(SD))2.2(0.56),2.8(0.80)和3.2(0.87)L / min(p <0.001)。 RATT,自行车测功机和跑步机分别为168(9.5),179(7.9)和184(6.9)次/分钟(p<0.001)。 RATT与VO2peak和HRpeak的对比自行车测功机与RATT与跑步机之间显示出很强的相关性。VO2peak和VO2peak的重测可靠性和重复性很高HRpeak适用于所有设备。结果表明,RATT是有效而可靠的运动测试设备。有潜力RATT用于无法使用标准的严重受损受试者方式。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号