Operator Variability in Scan Positioning is a Major Component of HR-pQCT Precision Error and is Reduced by Standardized Training

机译:扫描定位中的操作员差异是HR-pQCT精度误差的主要组成部分并且通过标准化培训来降低

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

IntroductionHR-pQCT is increasingly used to assess bone quality, fracture risk and anti-fracture interventions. The contribution of the operator has not been adequately accounted in measurement precision. Operators acquire a 2D projection (“scout view image”) and define the region to be scanned by positioning a “reference line” on a standard anatomical landmark. In this study, we (i) evaluated the contribution of positioning variability to in vivo measurement precision, (ii) measured intra- and inter-operator positioning variability, and (iii) tested if custom training software led to superior reproducibility in new operators compared to experienced operators.
机译:简介HR-pQCT越来越多地用于评估骨质量,骨折风险和抗骨折干预措施。在测量精度方面,运营商的贡献尚未得到充分考虑。操作员获取2D投影(“侦查视图图像”),并通过在标准解剖界标上放置“参考线”来定义要扫描的区域。在这项研究中,我们(i)评估了位置变异性对体内测量精度的贡献,(ii)测量了操作员内部和操作员之间的位置变异性,以及(iii)测试了定制培训软件是否比新操作员具有更高的可重复性给经验丰富的操作员。